A-19
Alarm limit
Range (mmHg)
Step (mmHg)
Sys High
Mean High
Dia High
(low limit + 2) to 300
Sys Low
Mean Low
ART
Ao
FAP
BAP
UAP
Dia Low
0 to (high limit – 2)
1
Sys High
Mean High
Dia High
(low limit + 2) to 120
Sys Low
Mean Low
PA
Dia Low
-6 to (high limit – 2)
1
Mean High
(low limit + 2) to 40
CVP, LAP
RAP, ICP,
UVP
Mean Low
-10 to (high limit – 2)
1
Sys High
Mean High
Dia High
(low limit + 2) to 300
Sys Low
Mean Low
P1 to P4
Dia Low
-50 to (high limit – 2)
1
A.5.8 C.O.
Measurement method
Thermodilution method
C.O.:
0.1 to 20 L/min
TB:
23 to 43
°
C
Measurement range
TI:
0 to 27
°
C
Resolution
C.O.:
TB, TI:
0.1 L/min
0.1
°
C
Accuracy
C.O.:
TB, TI:
±5% or ±0.1 L /min, whichever is greater
±0.1
°
C (without sensor)
Alarm range
BT:
23 to 43
°
C
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...