16-9
16.7 Removing Exhaust Gases from the System
WARNING
z
Anesthetics: When using the Sidestream or Microstream CO
2
measurement on
patients who are receiving or have recently received anesthetics, connect the outlet
to a scavenging system, or to the anesthesia machine/ventilator, to avoid exposing
medical staff to anesthetics.
To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet
connector of the module.
16.8 Zeroing the Sensor
The zero calibration eliminates the effect of baseline drift during CO
2
measurement exerted
on the readings and therefore maintains the accuracy of the CO
2
measurements.
16.8.1 For Sidestream and Microstream CO
2
Modules
For sidestream and microstream CO
2
modules, a zero calibration is carried out automatically
when necessary. You can also start a manual zero calibration if necessary. To manually start a
zero calibration, select [
Start Zero Cal.
] in the [
CO
2
Setup
] menu. Disconnecting the patient
airway is not required when performing a zero calibration.
16.8.2 For Mainstream CO
2
Modules
For mainstream CO
2
modules, zero the sensor whenever:
1. A new adapter is used;
2. You reconnect the sensor to the module;
3. You see the message [
CO
2
Zero Required
].In this case, check the airway adapter for
any blockage, e.g. mucus, etc. If a blockage is detected, clear or replace the adapter.
To zero the sensor, follow this procedure:
1. Connect the sensor to the module.
2. In
the
[
CO
2
Setup
] menu, set the [
Operating Mode
] to [
Measure
]. The message [
CO
2
Sensor Warmup
] is displayed.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...