14-9
14.7 Zeroing the Transducer
To avoid inaccurate pressure readings, the monitor requires a valid zero. Zero the transducer
in accordance with your hospital policy (at least once per day).Zero whenever:
A new transducer or adapter cable is used.
You reconnect the transducer cable to the monitor.
The monitor restarts.
You doubt the readings.
1. Turn off the stopcock to the patient.
2. Vent the transducer to the atmospheric pressure by turning on the stopcock to the air.
3. In the setup menu for the pressure (e.g. Art), select [
Zero Art >>
]
→
[
Zero
]. During
zero calibration, the [
Zero
] button appears dimmed. It recovers after the zero
calibration is completed.
4. After the zero calibration is completed, close the stopcock to the air and open the
stopcock to the patient.
NOTE
z
Your hospital policy may recommend that the ICP transducer is zeroed less
frequently than other transducers.
3-way stopcock
Pressure transducer
Pressure adapter cable
Monitor
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...