7
16.2.2 Using a Microstream CO
2
module ................................................................. 16-3
16.2.3 Using a Mainstream CO
2
module................................................................... 16-4
16.3 Changing CO
2
Settings ............................................................................................... 16-5
16.3.1 Entering the Standby Mode............................................................................ 16-5
16.3.2 Setting the Pressure Unit................................................................................ 16-5
16.3.3 Setting up Gas Compensations....................................................................... 16-5
16.3.4 Setting up Humidity Compensation ............................................................... 16-6
16.3.5 Setting the Apnea Alarm Delay...................................................................... 16-7
16.3.6 Choosing a Time Interval for Peak-Picking ................................................... 16-7
16.3.7 Setting the Flow Rate ..................................................................................... 16-7
16.3.8 Setting up the CO
2
Wave................................................................................ 16-7
16.4 Setting Barometric Pressure Compensation ................................................................ 16-8
16.5 Measurement Limitations............................................................................................ 16-8
16.6 Troubleshooting the Sidestream CO
2
Sampling System ............................................. 16-8
16.7 Removing Exhaust Gases from the System ................................................................ 16-9
16.8 Zeroing the Sensor ...................................................................................................... 16-9
16.8.1 For Sidestream and Microstream CO
2
Modules............................................. 16-9
16.8.2 For Mainstream CO
2
Modules ....................................................................... 16-9
16.9 Calibrating the Sensor ............................................................................................... 16-10
16.10 Oridion Information ................................................................................................ 16-10
17 Monitoring AG ............................................................................................................. 17-1
17.1 Introduction................................................................................................................. 17-1
17.2 Preparing to Measure AG............................................................................................ 17-2
17.3 Understanding the AG Display ................................................................................... 17-4
17.4 MAC Values................................................................................................................ 17-5
17.5 Changing AG Settings................................................................................................. 17-6
17.5.1 Selecting an Anesthetic Gas for Monitoring .................................................. 17-6
17.5.2 Setting Gas Unit ............................................................................................. 17-6
17.5.3 Setting the Apnea Alarm Delay...................................................................... 17-6
17.5.4 Changing the Sample Flow Rate.................................................................... 17-6
17.5.5 Setting up the O
2
Compensation .................................................................... 17-6
17.5.6 Entering the Standby Mode............................................................................ 17-7
17.5.7 Setting up the AG Wave ................................................................................. 17-7
17.6 Changing the Anesthetic Agent................................................................................... 17-7
17.7 Measurement Limitations............................................................................................ 17-8
17.8 Troubleshooting .......................................................................................................... 17-8
17.8.1 When the Gas Inlet is Blocked ....................................................................... 17-8
17.8.2 When an Internal Occlusion Occurs............................................................... 17-8
17.9 Removing Exhaust Gases from the System ................................................................ 17-9
18 Freezing Waveforms .................................................................................................... 18-1
18.1 Freezing Waveforms ................................................................................................... 18-1
18.2 Viewing Frozen Waveforms ........................................................................................ 18-1
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...