17-6
17.5 Changing AG Settings
17.5.1 Selecting an Anesthetic Gas for Monitoring
As the AION 02-type module cannot automatically identify 5 anesthetic gases, you need to
select an anesthetic gas before use. To select the anesthetic gas, in the [
Gas Setup
] menu,
select [
Agent
] and then select the appropriate setting.
17.5.2 Setting Gas Unit
For N
2
O and AA, the unit of the measured gas is fixed to “%”.
For EtCO
2
, select [
CO
2
Unit
] from the [
CO
2
Setup
] menu and toggle between [
mmHg
], [
%
]
and [
kPa
].
For AG module integrating O
2
module, you can set [
O
2
unit
] in the [
Gas Setup
] menu to
[
mmHg
], [
%
] or [
kPa
].
17.5.3 Setting the Apnea Alarm Delay
In the setup menu for any gas, select [
Apnea Time
] and select the appropriate setting. The
monitor will alarm if the patient has stopped breathing for longer than the preset apnea time.
17.5.4 Changing the Sample Flow Rate
In the setup menu for any gas, select [
Flow Rate
] and then choose:
[
200 ml/min
], [
150 ml/min
] or [
120 ml/min
] for adult and pediatric patients;
[
120 ml/min
], [
90 ml/min
] or [
70 ml/min
] for neonatal patients.
17.5.5 Setting up the O
2
Compensation
If the AG module does not incorporate the O
2
module, you need to manually select [
O2
Compen
] and then select [
Off
] or an appropriate setting according to the amount of O
2
in the
ventilation gas mixture. When the amount of O
2
is less than 30%, you’d better switch this
compensation off.
If the AG module incorporates the O
2
module, the system will directly use the O
2
concentration detected by the O
2
module to make compensation. At this time, the [
O2
Compen
] in the setup menu for any gas is fixed to [
Off
].
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...