15-3
3. Check if the height and weight are appropriate for your patient in the [
Patient
Demographics
] menu. Change if necessary.
4. Select the C.O. parameter window to enter the [
C.O. Setup
] menu. In the [
C.O. Setup
]
menu:
Check that the correct computation constant is entered. To change the
computation constant, select [
Comp. Const
] and then enter the correct value.
When a new catheter is used, the computation constant should be adjusted in
accordance with the manufacturer’s instructions for use.
Switch [
Auto Ti
] on and the injectate temperature is automatically obtained.
When [
Auto Ti
] is switched off, you can enter the injectate temperature in the
[
Manual Ti
] field.
5. Select
[
C.O. Measure
] to enter the C.O. measurements window.
Patient monitor
C.O. cable
Injectate
Syringe
In-line probe
Balloon inflation port
PA catheter
Ice water
Hole
Stopcock
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...