A-9
Pace Pulse
Pace pulses meeting the following conditions are labelled with a
PACE marker:
Pace pulse markers
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
When tested in accordance with the ANSI/AAMI EC13:
Sections 4.1.4.1 and 4.1.4.3, the heart rate meter rejects all
pulses meeting the following conditions.
Pace pulse rejection
Amplitude:
Width:
Rise time:
Minimum input slew rate:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
50 V/s RTI
Alarm limit
Range (bpm)
Step (bpm)
HR High
Adult:
Pediatric, Neonate:
(low limit+2) to 300
(low limit+2) to 350
HR Low
15 to (high limit-2)
1
Basic algorithm
HR
3-, 5-, and 12-lead ECG
Neonate:
Pediatric:
Adult:
15 to 350 bpm
15 to 350 bpm
15 to 300 bpm
Measurement range
Interpretation of resting
12-lead ECG
Adult:
30 to 120 bpm
Resolution 1
bpm
Accuracy
3-, 5-, and 12-lead ECG: ±1 bpm or ±1%, whichever is greater.
Interpretation of resting 12-lead ECG: ±2 bpm
Sensitivity 200
μ
V (lead II)
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...