D-7
Measurement
Alarm message
L
I
Cause and solution
CO2 Check Sensor
L
C
The CO2 sensor appears abnormal. Check or
replace the sensor.
CO2 No Sensor
L
A
Make sure that the CO2 sensor is connected.
CO2 Temp
Overrange
L
C
The module detects that the temperature has
exceeded the specified range. Please move away
the source of heat or wait for the module to
recover.
AG No Watertrap
L
B
Check the connections of the watertrap and
re-connect it.
AG Change
Watertrap
L
C
Wait until the change is completed.
AG Watertrap Type
Wrong
L
C
Make sure that a correct watertrap has been used.
YY Accuracy
Unspecified
(YY represents O2,
N2O, CO2, Enf, Iso,
Sev, Hal, Des, or
awRR (AG)
L
C
The measured value has exceeded the specified
accuracy range.
AG Hardware
Mulfunction
H
C
Stop using the AG module and contact your
service personnel.
AG Airway
Occluded
L
C
Check the airway and remove the occlusion.
AG Zero Failed
L
C
Restart the monitor, and then perform a zero
calibration again.
AG Cal. Failed
L
C
Re-calibrate the AG module.
Anesthetic Mixture
L
C
The system has detected more than one anesthesia
gas. Check the ventilated anesthesia gas.
O2 Sensor Err
L
C
The O2 sensor appears abnormal. Contact your
service personnel.
AG
AG Limit Error
L
C
AG violated set alarm limits seriously and
measurements cannot be performed. Contact your
service personnel.
12V Too High
H
C
12V Too Low
H
C
3.3V Too High
H
C
3.3V Too Low
H
C
There is a problem with the system power supply.
Restart the monitor.
Power
Low Battery
M C
Connect the monitor to an AC power source and
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...