7-9
7.5.5 Mass Alarm Setup
In the main menu, select [
Alarm Setup >>
]
→
[
Mass Alarm Setup>>
]. You can review and
set alarm limits, alarm switches and alarm recordings for all parameters.
7.5.6 Adjusting Alarm Limits Automatically
The monitor can automatically adjust alarm limits according to the measured vital signs,
using the auto limits function. When auto limits are selected, the monitor calculates safe
auto limits based on the latest measured values.
To get accurate auto alarm limits, you need to collect a set of measured vital signs as a
baseline. Then, in the main menu, select [
Alarm Setup >>
]
→
[
Mass Alarm
Setup>>
]
→
[
Auto Limits
]. The monitor will create new alarm limits based on the measured
values.
Before applying these automatically created alarm limits, confirm if they are appropriate for
your patient in the mass alarm setup menu. If not, you can adjust them manually. These
alarm limits will remain unchanged until you select auto limits again or adjust them
manually.
7.6 Latching Alarms
The alarm latching setting for your patient monitor defines how the alarm indicators behave
when you do not acknowledge them. When alarms are set to non-latching, their alarm
indications end when the alarm condition ends. If you switch alarm latching on, all visual
and audible alarm indications last until you acknowledge the alarms, except that the
measurement numeric and violated alarm limit stop flashing as soon as the alarm condition
goes away.
You can switch alarm latching on or off in configuration mode. For details, refer to the
chapter
Managing Configuration
.
You can clear the latched alarms by pressing the
hardkey.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...