8-22
8.8.2 Interpretation of resting 12-Lead ECG
Interpretation of resting 12-lead ECG is intended for adult patients.
You can only start an interpretation of resting 12-lead ECG 11 seconds after entering the
12-lead ECG monitoring screen. Otherwise, the prompt message [
Not enough data.
Cannot analyze.
] will be displayed. To start an interpretation of resting 12-lead ECG, select
[
Freeze
] and then [
Analyze
]. Then, the following screen will be displayed. In this screen,
you can:
Select [
Record Result
] to print out the interpretation of resting 12-lead ECG results by
the recorder.
Select [
Record Wave
] to print out the interpretation of resting 12-lead ECG results
and waves by the recorder.
Select [
Print Report
] to print out the interpretation of resting 12-lead ECG report by
the printer.
Besides, after selecting [
Freeze
], you can:
Browse the frozen ECG waves by selecting [
Scroll
] and rotating the Knob, or selecting
the or button beside [
Scroll
].
Print out the currently frozen waves by selecting [
Record
].
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...