A-5
A.2.5 Audio Indicator
Speaker
Give alarm tones (45 to 85 dB), key tones, QRS tones; support
PITCH TONE and multi-level tone modulation; alarm tones
comply with IEC60601-1-8.
A.2.6 Alarm Signal
Alarm output delay
Less than 1 second
Pause duration
1, 2, 3, 5, 10, 15 minutes or infinite, depending on configuration
A.2.7 Monitor Interface Specifications
Power
1 AC power input connector
Measurements
ECG, RESP, TEMP, SpO2, NIBP, IBP, CO, CO2, AG
Wire network
1 RJ45 connector, 100 Base-TX, IEEE 802.3
USB (optional)
2 connectors, USB 1.1
CF
50-pin CF revision 2.0 connector
Video interface
1 standard colour VGA monitor connector
Auxiliary output
1 connector, standard BNC, the common connector of nurse call
signals, analog output signals and defibrillator synchronization
signals
Equipotential Grounding
Terminal
1
RS232 connector
1 connector, supports DIAP communication protocol
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...