TOS6200A/6210
REMOTE and SIGNAL I/O
5-5
Table5-1
SIGNAL I/O Pin Configuration
No.
Signal name
I/O
Description of signal
1
PM0
I
LSB
LSD
2-digit BCD active-low input
Panel memory or program select signal input termi-
nal.
This select signal is latched at the rise of a strobe
signal to recall the panel memory or program.
2
PM1
I
3
PM2
I
4
PM3
I
5
PM4
I
MSD
6
PM5
I
7
PM6
I
8
PM7
I
MSB
9
STB
I
Panel memory or program strobe signal input terminal.
10
N.C
11
N.C
12
N.C
13
COM
Circuit common
14
TEST ON
O
ON during test.
15
TEST
O
ON during test (not including the period of a current/voltage rise.)
16
PASS
O
ON for approximately 0.2s or more when a measured value is judged
PASS. It is continuously ON when PASS HOLD is set to ON.
17
UPPER FAIL
O
Continuously ON if a measured value is the upper reference value or
greater and judged FAIL.
18
LOWER FAIL
O
Continuously ON if a measured value is the lower reference value or less
and judged FAIL.
19
READY
O
ON while the tester is in ready status.
20
PROTECTION
O
ON if PROTECTION status arises.
21
START
I
Start signal input terminal.
22
STOP
I
Stop signal input terminal.
23
ENABLE
I
Start signal’s ENABLE signal input terminal.
24
+24 V
O
+24V internal power output terminal. Maximum output current is 100
mA.
25
COM
Circuit common
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...