4-4
Basic Operation
TOS6200A/6210
4.2.1
Test Current
Set the maximum current value flowing through the DUT as follows:
TOS6200A: 3.0 A to 31.0 A
TOS6210: 6.0 A to 62.0 A (resolution: 0.1 A)
1.
Press the F1 (CURRENT) key to move the cursor to the test current
value if it is not already there (you can also move the cursor using the
▲▼
keys).
2.
Use the
keys to move the cursor below a digit to be set.
3.
Set a test current value with the rotary knob.
“OVER VOLT” blinks on the LCD.
A value obtained by multiplying the test current value by the upper reference
value has exceeded 5.4 V. Test cannot begin under the current conditions.
Change the setting referring to “OVER VOLT” in 4.9.1, "Setting Output out of
the Operation Range.”
“OVER VA” blinks on the LCD.
The setting has exceeded the maximum rated output. Test cannot begin under
the current conditions. Change the setting referring to “OVER VA” in 4.9.1,
"Setting Output out of the Operation Range.”
“OVER RESI” blinks on the LCD.
A value obtained by dividing the upper reference value by the test current value
has exceeded 0.6
. Test cannot begin under the current conditions. Change the
setting referring to “OVER RESI” in 4.9.1, "Setting Output out of the Operation
Range.”
The current value may be changed even during testing.
To avoid rapid changes, current value must be changed by increasing or decreas-
ing the least significant digit (the cursor is fixed at the least significant digit).
If the test current is changed when the resistance value is close to the upper ref-
erence value, a FAIL status may occur.
“OVER LOAD” blinks on the LCD during testing.
If the output power exceeds a maximum power rating, the power limitation
safety function is tripped and protection status occurs.
Change the setting referring to 4.10.3, "Overload Protection (OVER LOAD).”
CURRENT LOWER UPPER
TIMER
25.0
A
0.100 60.0
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
READY
50Hz
UPPER
F1 F2 F3 F4 F5
▲
▲
▲
▲
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...