4.10.4 Output Voltage Limitation (VOLT LIMIT)
This protective function limits the output voltage. If the output terminal voltage
exceeds 5.6 V when the measured value is not judged FAIL during a test, the output
voltage limitation function is tripped and the "VOLT LIMIT" indication is blinked
on the LCD.
Since no test can be performed for settings for which the product of the test current
and upper reference value (including offset value) exceeds 5.4 V, the output terminal
voltage will generally not exceed 5.6 V and the output voltage limitation function is
rarely tripped. However, for measurements using four terminals that use high-resis-
tance lead wires, or for two-terminal measurements with the offset canceling func-
tion activated, the output terminal voltage may exceed 5.6 V in some cases, tripping
this function. This causes the tester to enter protection status.
For example, assume that the DUT with 0.190
is tested with the test current set to
25 A and upper reference value set to 0.200
. If the leads connected to the DUT
are long and four-terminal measurements are made using leads exceeding 0.034
both coming and going, the combined resistance from the perspective of the output
terminals becomes 0.224
or greater, tripping the output voltage limit function for
5.6 V and disabling the test.
If the product of the test current and a combined resistance from the perspective of
the output terminals is closer to 5.6 V, use leads with lower resistance values.
Press the STOP switch to cancel output voltage limitation.
CURRENT LOWER UPPER
TIMER
25.0
A
0.200 60.0
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
50Hz
UPPER
VOLT LIMIT
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...