6-26
GPIB and RS-232C
TOS6200A/6210
6.3.4
Messages Relating to the Tester Status
This subsection describes device messages used to check the status of the tester or
measured values.
FUNCTION (FUN)
Shifts the LCD to each mode (setting screen) or inquires about the current mode
(displayed screen).
Message
Command message:
Fun{0|1|2|3|4}
Query message:
FUN?
Program data
Set value:
0
Test conditions setup screen (MAIN)
1
Program screen (AUTO)
2
Program edit screen (AUTO EDIT)
3
System setup screen (SYSTEM1)
4
Offset screen (OFFSET)
5
Those other than 1 to 4 above (not avail-
able)
Response
Returns the current mode (displayed screen) in response to FUN?.
(Example) When the current mode (displayed screen) is any figure other than 0 to 4
A value of
5
is returned.
IDATA? (IDAT?)
Inquires about the present measured current value.
Message
Query message:
IDAT?
Response
Returns the present measured current value in response to IDAT?
It returns the current value being measured during a test or the measured current
value obtained immediately before when the test was completed.
(Example) When the present measured current value is 25.0 A
A value of
25.0
is returned.
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...