TOS6200A/6210
GPIB and RS-232C
6-39
Fail Register
Invalid Setting Register
Error Register
Bit
Descriiption
7
Not used in the tester
6
Not used in the tester
5
Not used in the tester
4
Not used in the tester
3
Not used in the tester
2
UPPER FAIL
Indicates that the result has been judged as FAIL, exceeding the
upper reference limit.
1
LOWER FAIL
Indicates that the result has been judged as FAIL, exceeding the
lower reference limit.
0
Not used in the tester
Bit
Descriiption
7
Not used in the tester
6
Not used in the tester
5
Not used in the tester
4
Not used in the tester
3
OVER RESI(Over Resistance)
Indicates that the setting exceeds the resistance measurement
range.
2
OVER VA(Over Volt-Amp)
Indicates that setting exceeds the power limit.
1
UP
LOW
Indicates that the set value for lower limit judgement exceeds
the upper reference value.
0
OVER V(Over Voltage)
Indicates that the test voltage has exceeded 5.4 V.
Bit
Descriiption
7
Not used in the tester
6
Not used in the tester
5
Not used in the tester
4
Not used in the tester
3
Invalid device message
Indicates that an invalid device message was generated.
2
Out-of-range error
Indicates occurrence of out-of-range error.
1
Data error
Indicates occurrence of data error.
0
Syntax error
Indicates occurrence of header error.
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...