background image

1-4

General

TOS6200A/6210

Reduced cycle time

A test current is a constant current value set within approximately 100 ms.

Earth continuity tests can be conducted even at 1-second intervals. This enables
the instrument to perform tests in production lines that require reduced cycle
times.

Compact and lightweight

A new high-efficiency power supply (achieving a power conversion efficiency of

65%) and large output (TOS6200A: 150 VA, TOS6210: 220 VA) makes possible a
tester that is remarkably compact and lightweight, about half the size and weight
of our conventional testers.

Excellent measurement accuracy

The tester is equipped with an ohmmeter of +/-(2% of r 0.003

) that cal-

culates resistance values based on measured current and voltage.

Offset canceling function

The tester has an offset canceling function that cancels values for contact resis-
tance at alligator clip connections, the contact resistance of measuring leads in
measurements using two terminals, and other resistance components.

Contact check function

The tester is provided with a contact check function that verifies the connection of
the DUT (by current detection) before initiating testing.

Voltage judgment function

This function allows either a resistance value or a voltage drop value (voltage
between the SAMPLING terminals) to be selected as the reference value for judg-
ment criteria.

It also allows voltage drop value-based testing to be conducted, which is required
by safety standards such as UL 60950-1 and IEC 60950-1.

Simple operations

Tester functions are fast and easily controlled, with an intuitive control method
that's easy to pick up, even for first-time users.

For example, test conditions are set simply by selecting an item displayed on the
LCD with the cursor keys and turning the rotary knob to set/select a value.

Any items you wish to set up can be assigned to the function keys.

CH1: Current waveform

CH1: 50 A/div 

 

 

20 ms/div

TOS6210

Содержание TOS6200A

Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...

Страница 2: ...ed or missing pages in this manual it will be replaced If the manual gets lost or soiled a new copy can be provided for a fee In either case please contact your Kikusui agent or distribu tor and provi...

Страница 3: ...ltage is output here DANGER Indicates an imminently hazardous situation which if ignored will result in death or serious injury Indicates a potentially hazardous situation which if ignored could resul...

Страница 4: ...ing Four Terminals 2 9 2 4 5 Measurements Using Two Terminals 2 9 2 4 6 Connecting to the DUT 2 10 2 5 Preliminary Inspection 2 11 Chapter 3 Part Names and Functions _________________________ 3 1 3 1...

Страница 5: ...hange in ENABLE Signal SIGNAL I O 4 40 4 10 6 Output current error CURR ERROR 4 40 4 11 Initialize 4 41 Chapter 5 REMOTE and SIGNAL I O _________________________ 5 1 5 1 REMOTE Terminal 5 2 5 2 SIGNAL...

Страница 6: ...se of Problems 7 5 Chapter 8 Specifications ___________________________________ 8 1 8 1 Basic Performance 8 2 8 2 Interface and Other Functions 8 4 8 3 General Specifications 8 6 8 4 External Dimensio...

Страница 7: ...TOS6200A 6210 1 1 Chapter 1 General Gives an overview of the tester and describes its features and various options 1 1...

Страница 8: ...g screen displayed immediately after power is switched ON You can also obtain the ROM version number with the IDN message For information on the IDN message see 6 3 1 Register Related and General Purp...

Страница 9: ...test changes Safety output voltage The constant current constant voltage circuit provides fast response preventing generation of excessive output voltage even when output is interrupted during test in...

Страница 10: ...gator clip connections the contact resistance of measuring leads in measurements using two terminals and other resistance components Contact check function The tester is provided with a contact check...

Страница 11: ...he remote control Programming of test conditions A combination of stored test conditions allows automatic execution of tests con sisting of several programs of up to 100 steps each Although the total...

Страница 12: ...test starts only when both switches are pressed simultaneously Functions OPERATE switch Test switches are enabled only when this switch is set to the ON position When it is set to the OFF position a...

Страница 13: ...l tester starting test stopping test operations Connect it to the OUTPUT and REMOTE terminals on the tester s front panel Maximum rating LP01 TOS 30 A LP02 TOS 60 A Cable length 2 m Accessory LOW Test...

Страница 14: ...1 8 General TOS6200A 6210...

Страница 15: ...TOS6200A 6210 2 1 Chapter 2 Installation and Preparation for Use Describes the steps from unpacking to installation to preparation required before switching on POWER 2 2...

Страница 16: ...cking materials may be used for later transport of the product so it is recom mended that they be retained Rated voltage 250 Vac PLUG CEE7 7 85 AA 0005 Rated voltage 250 Vac PLUG GB1002 85 10 0790 or...

Страница 17: ...0 rh no dew condensation is allowed When you store the product be sure to observe the temperature and humidity ranges indicated below Storage temperature range 20 C to 70 C 4 F to 158 F Storage humidi...

Страница 18: ...th a plug can be used to disconnect the product from the AC line in an emergency Connect the plug to an easily accessible power outlet so that the plug can be removed from the outlet at any time Be su...

Страница 19: ...rge current and consequently generates a strong magnetic field Make sure that no articles that may be affected by mag netic fields are located near the test leads or current output lines For example i...

Страница 20: ...ult in inaccurate measurements and burns or injury resulting from heat generated by contact resistance at the terminals 4 Connect the alligator clip of the test lead black or white to the protec tive...

Страница 21: ...he rated current Improper terminal connections can result in inaccurate measurements and burns or injury resulting from heat generated by contact resistance at the terminals 1 Turn off the POWER switc...

Страница 22: ...ermi nals appropriate for the sectional area of the current output line Select wires on the basis of the following criteria Resistance values with respect to the nominal sectional areas of wires are a...

Страница 23: ...the OUTPUT terminals sub tracted beforehand For more information see 4 2 6 Offset Canceling Function OUTPUT LOW HIGH LOW HIGH SAMPLING r1 r3 r5 r6 r7 r8 r4 r2 R2 R3 R4 DUT Device Under Test r1 to r8...

Страница 24: ...the test probe or lead during the test or immediately after test since they are at high temperature Testing from the protective conductor terminal of the AC power inlet Connect one of the test leads t...

Страница 25: ...power cord protective ground wire from testing Check the appropriate safety standard to determine if this is the case 2 5 Preliminary Inspection Always inspect the following four points before testin...

Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...

Страница 27: ...TOS6200A 6210 3 1 Chapter 3 Part Names and Functions Gives the names and functions of switches keys indications connectors and other parts on the front and rear panels 3 3...

Страница 28: ...witch places the tester in ready status 3 START This switch is used to start a test Press this switch when READY is displayed on the LCD to begin testing During testing the TEST LED indicator lights a...

Страница 29: ...g this limit 6 LCD Displays information including the range of set values and measured values 7 Feet Used to raise the front panel of the tester to get a better view of the LCD screen or improve the o...

Страница 30: ...splays the offset measurement screen OFFSET 12 MEMORY RECALL STORE key Press this key to recall panel memory Change the memory number using the rotary knob then press the ENTER key next to this key Th...

Страница 31: ...NEXT 16 Indicators TEST This LED indicates that testing is underway Blinks for contact checks when the contact check is set to ON standby PASS This LED gives the test result When the test result has b...

Страница 32: ...of the tester through a PCvia the GPIB interface 24 RS 232C This connector connects an RS 232C cable for remote control of the tester is remote through a PC via the RS 232C interface 25 Fan Used to co...

Страница 33: ...ed to wind the AC power cord when the tester is not in use Never use the tester in an upright position Use of the product with the cord holders serving as feet causes the tester to fall down which may...

Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...

Страница 35: ...TOS6200A 6210 4 1 Chapter 4 Basic Operation Describes basic operations such as setting test conditions and starting a test 4 4...

Страница 36: ...the LCD the screen displayed when you last turned the POWER switch off appears When the switch is turned on for the first time after pur chase the following test conditions setup screen MAIN is displa...

Страница 37: ...to a desired item Holding down with the SHIFT key press any of keys F1 to F5 to activate the operation indicated above the key Settings corresponding to safety standards At factory shipment settings...

Страница 38: ...etting referring to OVER VA in 4 9 1 Setting Output out of the Operation Range OVER RESI blinks on the LCD A value obtained by dividing the upper reference value by the test current value has exceeded...

Страница 39: ...Select resistance value based judgment or sampled voltage value based judgment Press the SHIFT and F4 keys together to switch between judgment methods Each time you press the SHIFT and F4 keys the up...

Страница 40: ...ions In this case increase the upper reference value or reduce the lower reference value OVER VOLT blinks on the LCD A value obtained by multiplying the test current value by the upper reference value...

Страница 41: ...se increase the upper reference value or reduce the lower reference value OVER VA blinks on the LCD The setting has exceeded the maximum rated output Test cannot begin under the current conditions Cha...

Страница 42: ...t Turning the knob clockwise ON Turning the knob counterclockwise OFF Selection of the judgment method Select resistance value based judgment or sampled voltage value based judgment Press the SHIFT an...

Страница 43: ...s to move the cursor to the desired digit 3 Use the rotary knob to set a lower reference value UP LOW blinks on the LCD If a lower reference value more than the upper reference value is set when lower...

Страница 44: ...rsor to the desired digit 3 Use the rotary knob to set a lower reference value UP LOW blinks on the LCD If a lower reference value more than the upper reference value is set when lower limit judgment...

Страница 45: ...F5 toggles between ON and OFF You can also use the keys to move the cursor to the timer function ON OFF indicator With the cursor at the TIMER indication use the rotary knob to set Turning the knob cl...

Страница 46: ...onent offset value in advance using this tester Offset Canceling Function ON OFF Settings To use the offset canceling function set the OFFSET indication to ON Pressing the SHIFT and F3 keys together a...

Страница 47: ...et measurement ends the tester reverts to ready status Press the MAIN key to display the MAIN screen Check that the offset function is ON then begin the test To check the offset value display the OFFS...

Страница 48: ...tion on MOMENTARY see 4 4 System Setup 4 3 1 Starting a Test To begin the test To begin testing press the START switch while the MAIN screen indicates ready In the ready status the upper right part of...

Страница 49: ...switch was pressed The tester has entered protection status PASS judgment When the lower limit judgment function is ON and the measured value falls between the upper and lower reference values the te...

Страница 50: ...nge between the upper and lower reference values the tester returns a FAIL judg ment When the lower limit judgment function is OFF the tester returns a FAIL for the measured value if the value exceeds...

Страница 51: ...RM Normal measurement mode MAX Retains a peak value 1 Press the F1 MEAS key or use the keys to move the cursor to MEAS MODE 2 Use the rotary knob to select either MAX or NORM Turning the knob clockwis...

Страница 52: ...has been set to HOLD for pro grammed automatic testing the program cannot proceed to the next step ON OFF for Fail Mode With FAIL MODE set to ON FAIL judgment and protection status cannot be can cell...

Страница 53: ...erminals When you press the START switch with the OUTPUT terminals opened a TEST indication appears on the LCD and the TEST LED on the front panel blinks and the tester enters current monitoring statu...

Страница 54: ...rsor to BUZZER VOL Press the F1 BUZZER key to hear the buzzer sound set 3 Use the rotary knob to set the buzzer volume Setting the Contrast You can set the LCD screen contrast in the range from 0 to 1...

Страница 55: ...ee Appendix A 2 1 If COMMENT is not indicated on the LCD press the F5 NEXT function key to display the SYSTEM3 screen The display can also be switched by pressing the SHIFT and keys together 2 Use the...

Страница 56: ...TERFACE screen GPIB address RS 232C protocol Communication rate Data length Parity bit Stop bit 4 5 1 GPIB Address Set the GPIB address of the tester in the range from 0 to 30 The GPIB address set is...

Страница 57: ...the following two choices 7 bit 8 bit 1 Press the F3 DATA key or use the keys to move the cursor to DATA 2 Use the rotary knob to select either 7 or 8 Setting the Parity Bit Select a parity bit from t...

Страница 58: ...F setting Upper reference value Timer including ON OFF setting Measured offset value including ON OFF setting On shipment from the factory settings corresponding to a variety of safety standards are w...

Страница 59: ...ht of the screen title 3 Use the rotary knob to set the memory number from 00 to 99 to which test conditions are to be stored 4 Press the key to move the cursor to UNTITLED 5 Use the rotary knob to en...

Страница 60: ...99 3 Press the ENTER key This recalls the test conditions for that memory number Moving the cursor to another area of the screen before pressing the ENTER key aborts the recall operation Modifying the...

Страница 61: ...performs a test of memory 01 in step 00 and then that of memory 02 in step 01 9 9 seconds after that Because the interval time of step 01 is set to HOLD the test of memory 3 in step 02 does not start...

Страница 62: ...r step setting Interval time Repeat setting Setting a Program Name Enter the name of a program to create up to 12 characters as follows 1 Press the SHIFT F1 key combination pressing the F1 key with th...

Страница 63: ...s the F1 INS key to add the step MEM00 1 0s is added to the end of the list 4 Press the key to move the cursor to MEM to the right of the step number 5 Use the rotary knob to set a memory number 6 Use...

Страница 64: ...the upper right corner of the LCD and the TEST LED indicator lights up A running step will be displayed in reverse video 4 7 4 Suspending the Program Press the STOP switch to halt testing during progr...

Страница 65: ...OS6200A 6210 Basic Operation 4 31 4 7 6 Ending the Program To exit the program mode and return to normal status press the MAIN key This turns off the LED for the AUTO key and redisplays the MAIN scree...

Страница 66: ...the SHIFT LOCAL KEY LOCK keys again 4 9 Checking the Test Conditions This tester checks if the values set as the test conditions are valid or not If an invalid value is set the tester displays the ca...

Страница 67: ...Aac TOS6210 6 0 Aac to 62 0 Aac The following figures show the operation range of the testers If a test condition out of this range is set the LCD displays the cause of it a b c OVER RESI OVER VOLT OV...

Страница 68: ...setting exceeding the operation range is made the READY indication in the upper right of the LCD changes to blink OVER VA informing you that testing cannot be conducted under the current conditions TO...

Страница 69: ...ithin the operation range If a resistance value is set to the upper reference value OVER RESI check is not made 4 9 2 Upper Reference Value Lower Reference Value UP LOW If a lower reference value more...

Страница 70: ...ECTION LED indicator lights and the test is halted The LCD displays the cause of bringing the tester to protection status LCD Indication Given in Protection Status Cause of Bringing the Tester to Prot...

Страница 71: ...n blinks on the LCD stop testing for approximately the pause time specified in the table below and then press the STOP key If the output section temperature has returned to the nor mal level the teste...

Страница 72: ...value is 5 4 V or less and a test can be performed the overload protection function is tripped if the output power exceeds the following maximum rated output during testing This causes the OVER LOAD i...

Страница 73: ...ffset canceling func tion activated the output terminal voltage may exceed 5 6 V in some cases tripping this function This causes the tester to enter protection status For example assume that the DUT...

Страница 74: ...rent when the measured value is not judged FAIL during a test the output current error function is tripped and the CURR ERROR indication is blinked on the LCD If the output terminal voltage exceeds 5...

Страница 75: ...ress the SHIFT key until the charac ters KIKUSUI ELECTRONICS CORP disappear Factory shipped settings TOS6200A TOS6210 Test conditions set up screen MAIN Test current CURRENT 3 0 A 6 0 A Frequency FREQ...

Страница 76: ...4 42 Basic Operation TOS6200A 6210...

Страница 77: ...TOS6200A 6210 5 1 Chapter 5 REMOTE and SIGNAL I O Describes the REMOTE terminal for connecting optional equipment and the SIGNAL I O connector for input output signals 5 5...

Страница 78: ...ctor located on the front panel This connector is used to connect the optional RC01 TOS or RC02 TOS remote control boxes or the LP01 TOS or LP02 TOS test probe When you connect an option the START and...

Страница 79: ...2 D SUB 25 pin female connector Screw M2 6 x 0 45 Complies connector plug D sub 25 pin male with fix screw M2 6 To avoid malfunctions caused by noise use shielded type D sub 25 pin connectors and a ca...

Страница 80: ...withstand voltage 30 Vdc Output saturation voltage Approximately 1 1 V at 25 C Maximum output current 400 mA total Pin configuration Fig 5 3 SIGNAL I O Pin Configuration To avoid damage to internal ci...

Страница 81: ...g test not including the period of a current voltage rise 16 PASS O ON for approximately 0 2s or more when a measured value is judged PASS It is continuously ON when PASS HOLD is set to ON 17 UPPER FA...

Страница 82: ...er a lapse of 10 ms or more from the READY signal has turned to a low level set the START signal to a low level for 5 ms or more The READY signal turns to a high level after the effective START signal...

Страница 83: ...M Signals and Panel Memory or Program Number to be Recalled MSD LSD MAIN AUTO PM7 PM6 PM5 PM4 PM3 PM2 PM1 PM0 H H H H H H H H Recalls panel memory 0 Recalls program 0 H H H H H H H L Recalls panel mem...

Страница 84: ...t such as transistors instead of a switch in the above example Configure the circuit so that a collector current ic of 5 mA or greater passes through the transistor Output signal Example of driving a...

Страница 85: ...device messages used to create a program for remote control of the tester via GPIB or RS 232C To avoid electric shocks always turn off the power supply of the related equipment when connecting or disc...

Страница 86: ...ntact Kikusui distributor agent GPIB cable 1 m code 92080 GPIB cable 2 m code 92070 GPIB cable 4 m code 92090 6 1 2 RS 232C Interface To use the RS 232C interface make the following settings For setti...

Страница 87: ...l ready 5 GND Signal ground 6 DSR Data set ready 7 RTS Request to send 8 CTS Clear to send 9 RI Ring indicator Controller side 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 D sub pin 9 female D sub pin 9 female...

Страница 88: ...the controller to the device and responses sent from the device to the controller Each message consists of a message header and data sections Command Commands include settings and queries Setting Use...

Страница 89: ...owever in the event of a message terminator with only EOI use Acknowledgment messages RS 232C The acknowledgment message is peculiar to the RS 232C interface This message is sent from the tester to a...

Страница 90: ...ption A character string within these brackets represents program data Do not use these brackets when writing comments Characters or numbers separated by within these brackets indi cate that you can c...

Страница 91: ...ose device messages used to specify of a terminator and other items CLS Clears of the status byte register event status register device status register protec tion register fail register invalid setti...

Страница 92: ...to set or reset each bit of the service request enable register Also used to inquire about the contents of the service request enable register For details on the service request enable register see 6...

Страница 93: ...r SDC has been received Also used to perform the same processing as is performed by a DCL message from the RS 232C Message Command message CLR DSE Used to set or reset each bit of the device status en...

Страница 94: ...A value of 1 is returned Example When a test is being made A value of 12 is returned 4 Test 8 Test on Example When contact check is ON A value of 8 is returned ERR Used to inquire about the contents o...

Страница 95: ...alid setting register For details on the invalid setting register see 6 4 Registers Message Query message INV Response Returns the contents of the invalid setting register to INV Example If Bit 0 OVER...

Страница 96: ...r your PC Message Command message SIL 0 1 Query message SIL Program data Set value 0 Return an acknowledgement message 1 Do not return an acknowledgement mes sage default Response The set value for th...

Страница 97: ...on the tester SYS TEM screen BUZZERVOL BVOL Sets or inquires about the buzzer volume Message Command message BVOL NR1 Query message BVOL Program data Set value 1 to 10 default is 4 Resolution 1 Examp...

Страница 98: ...line has 20 characters then returns this as the result Example If KIKUSUI is entered in the 1st line no data is entered in the 2nd line and 123456789ABCDEFGHIJK is in the 3rd line the response messag...

Страница 99: ...in response to CON Example When contrast is set to 3 A value of 3 is returned DOUBLEACTION DAC Sets ON OFF for start double action or inquires about ON OFF status Message Command message DAC 0 1 OFF O...

Страница 100: ...OFF status of the fail mode in response to FMOD Example When fail mode is ON A value of 1 is returned MEASMODE MMOD Sets or inquires about the measurement mode Message Command message MMOD NORM MAX Q...

Страница 101: ...Momentary ON Response Returns the ON OFF status of the momentary in response to MOM Example When the momentary is ON A value of 1 is returned PASSHOLD PHOL Sets the PASS hold time or inquires about th...

Страница 102: ...under the current conditions Change the set ting referring to OVER VOLT in 4 9 1 Setting Output out of the Operation Range OVER VA blinks on the LCD The setting has exceeded the maximum rated output T...

Страница 103: ...urns the set value of test frequency in response to FREQ Example When the test frequency is 50 Hz A value of 50 is returned RJUDGE RJUD Sets ON OFF of resistance value based judgment or inquires about...

Страница 104: ...LOW Sets a resistance value based lower reference value and ON OFF of the lower limit judgment function or inquires about the ON OFF status of the lower limit judgment function When you perform sampl...

Страница 105: ...command causes sampled voltage value based judgment to be automatically selected Message Command message VLOW NR2 0 1 OFF ON Query message VLOW Program data NR2 Lower reference value Set value 0 01 to...

Страница 106: ...he upper right of the LCD changes to blink UP LOW informing you that testing cannot begin under the current conditions In this case increase the upper reference value or reduce the lower reference val...

Страница 107: ...LCD changes to blink UP LOW informing you that testing cannot begin under the current conditions In this case increase the upper reference value or reduce the lower reference value OVER VA blinks on...

Страница 108: ...test time and the timer ON OFF setting Message Command message TIM timer 0 1 OFF ON Query message TIM Program data timer NR2 Test time Set value 0 3 to 999 default is 1 0 Resolution 0 3 to 99 9 0 1 1...

Страница 109: ...due to a HOLD setting in the interval to the next step This command message is valid only in the test conditions set up MAIN offset measurement OFFSET or program AUTO screens Switch screens using the...

Страница 110: ...ystem setup screen SYSTEM1 4 Offset screen OFFSET 5 Those other than 1 to 4 above not avail able Response Returns the current mode displayed screen in response to FUN Example When the current mode dis...

Страница 111: ...the device status register TEST and TEST ON bit 2 3 are 1 measured voltage value 4 50 V measured current value 25 0 A maxi mum measured resistance value 0 182 measured resistance value 0 180 and elap...

Страница 112: ...d test time if the timer is OFF or the remaining time if the timer is ON Message Query message TIME Response Returns the elapsed or remaining time of the test in response to TIME Example When the time...

Страница 113: ...ge MEM memory no Program dataSets the memory contents in the following order All parameters must be entered even when the lower limit judgment is set to OFF Enter any value into the lower reference va...

Страница 114: ...Y VMEM has no query To inquire about the memory contents set using VMEM use MEMORY MEM In this case the voltage value is returned to the upper and lower refer ence values Messages Command message VMEM...

Страница 115: ...memory no Example When memory contents 10 is memory name TEST1 test current 62 0 A upper reference value 5 4 V lower reference value 0 05 V test time 999 s test frequency 50 Hz lower limit judgment O...

Страница 116: ...1 step number NR1 Set value 0 to 99 Resolution 1 Example To delete step 10 in program number 1 PDEL 1 10 PRGEDIT PED Overwrites a currently set step with a new step adds a step in a program or inquire...

Страница 117: ...se to PED program number step number Example When PED 5 1 inquires about the contents of step 1 in program 5 and step 1 is configured as memory 11 and an interval time of 5 seconds A value of the 11 5...

Страница 118: ...a name of TEST1 to program number 1 PNAM 1 TEST1 Response Returns the name of the program in response to PNAM program number Example When program 1 named TEST1 is inquired about with a PNAM 1 A value...

Страница 119: ...s returned PRGTEST PTES Recalls program contents Message Command message PTES NR1 Program data Set value 0 to 99 Resolution 1 PRGTOTAL PTOT Inquires about the number of steps corresponding to the prog...

Страница 120: ...DSR Read using DSE Read using ESR Status byte register 7 6 7 6 CME EXE 3 2 1 0 3 2 1 0 3 2 1 0 TEST ON TEST FAIL STOP PROT PASS READY INV SET Protection register Invalid setting register Read using P...

Страница 121: ...cats that a bit of the event status register has been set 4 DSB Device Status Bit Indicats that a bit of the device status register has been set 3 Not used in the tester 2 Not used in the tester 1 Not...

Страница 122: ...ADY Ready status Bit Descriiption 7 Not used in the tester 6 Not used in the tester 5 CURR Current Error When output current is more than 10 of the set value for current 4 REN Remote Enable Change in...

Страница 123: ...the tester 4 Not used in the tester 3 OVER RESI Over Resistance Indicates that the setting exceeds the resistance measurement range 2 OVER VA Over Volt Amp Indicates that setting exceeds the power li...

Страница 124: ...ZZERVOL BVOL 1 to 10 1 Setting for the buzzer volume 9 BUZZERVOL BVOL Returns the value set for the buzzer volume 10 CLR Clears the input buffer and has it wait for a device mes sage Equivalent to DCL...

Страница 125: ...t judg ment function 36 MEASMODE MMOD NORM MAX Setting for the measurement mode 37 MEASMODE MMOD Returns the setting status of the measured result indica tion 38 MEMORY MEM 0 to 99 1 Memory number 20H...

Страница 126: ...IT PED 0 to 99 1 Program number 0 to 99 1 Step number 0 to 99 1 Memory number 0 to 9 9 HOLD 0 1 s Interval time up to the next step 49 PRGEDIT PED 0 to 99 1 Program number 0 to 99 1 Step number Return...

Страница 127: ...TRM Returns a current response message terminator 73 UPPER UPP 0 001 to 1 200 TOS6200A 0 001 to 0 600 TOS6210 0 001 OHM Setting for upper reference value 74 UPPER UPP Returns the upper reference value...

Страница 128: ...and RS 232C TOS6200A 6210 81 VUPPER VUPP 0 01 to 5 40 0 01 V Sets the upper reference value 82 VUPPER VUPP Returns the upper reference value Header Data Function and response data 1 2 Set value Reso l...

Страница 129: ...6200A 6210 7 1 Chapter 7 Maintenance Explains tester maintenance and inspection of the tester To maintain high performance for an extended period perform maintenance and inspect the tester regularly 7...

Страница 130: ...discoloration or erase printed characters clouded display and so on 7 2 Inspection Any break or cracking in the covering greatly increases the risk of electric shock or fires If you find any breaks or...

Страница 131: ...fferent rating never short circuit the fuse holder 1 Turn off the POWER switch and unplug the AC power cord from the electrical outlet 2 Disconnect the AC power cord from the AC LINE connector on the...

Страница 132: ...service condi tions However we recommend replacing the fan roughly every 60 000 hours We also recommend regular internal inspections and cleaning at the same interval Backup battery The tester uses a...

Страница 133: ...SIGNAL I O 5 1 Check that the tester is not in PROTEC TION PASS or FAIL status 4 10 Protection Fuction 4 36 Check that the tester is not in the midst of system setup interface setup or pro gram editin...

Страница 134: ...7 6 Maintenance TOS6200A 6210...

Страница 135: ...nless otherwise specified the specifications contained in this chapter are based on the following settings and conditions Warm up time 30 minutes Temperature 5 C to 35 C Humidity 20 rh to 80 rh no con...

Страница 136: ...ent range 0 00 Vac to 6 00 Vac Resolution 0 01 V Offset cancel function 0 00 V to 5 40 V Offset ON OFF function provided Accuracy 1 of reading 0 02 V Response Mean value response rms value display res...

Страница 137: ...efer ence value UPPER 0 01 V to 5 40 V Setting range for the lower refer ence value LOWER 0 01 V to 5 40 V Resolution 0 01 V Judgment accuracy 2 of setting 0 05 V Calibration Calibration is performed...

Страница 138: ...on All functions except for the POWER switch and KEY LOCK function may be controlled remotely Baud rate 9600 19200 or 38400 bps GPIB Complies with IEEE Std 488 1978 Function All functions except for t...

Страница 139: ...nd judged FAIL 18 L FAIL O Continuously ON if a measured value equal to or below the lower reference value is detected and judged FAIL 19 READY O ON while the tester is in ready status 20 PROTECTION O...

Страница 140: ...of the following directive and standard EMC Directive 2004 108 EC EN 61326 1 Class A 7 EN 55011 Class A 7 Group 1 8 EN 61000 3 2 EN 61000 3 3 Under following conditions 1 Used test leadwire TL11 TOS f...

Страница 141: ...nsions Weight Approx 9 kg 19 84 lbs Approx 11 kg 24 25 lbs Accessories AC power cord 1 piece Test leadwire TL11 TOS 1 set TL12 TOS 1 set Short bar 2 pieces These are inserted between the OUTPUT and SA...

Страница 142: ...rnal dimensions Fig 8 2 TOS6210 external dimensions MAX345 13 58 35 1 4 MAX 270 10 63 MAX455 17 91 MAX10 0 39 430 16 93 MAX15 0 59 88 3 46 MAX140 5 51 Unit mm inch 430 16 93 MAX 10 0 39 MAX455 17 91 8...

Страница 143: ...se to output This compact lightweight PWM sine wave inverter is capable of a maximum output of 150 VA TOS6200A 220 VA TOS6210 achieving a power conversion effi ciency of 90 3 Output transformer This s...

Страница 144: ...ge control circuit Sends the detected output voltage to the power amplifier circuit for constant voltage at the output terminals It also DC converts the detected output voltage and displays averaged d...

Страница 145: ...3 IEC60065 1 25 A 0 1 0 001 60 s 50 Hz 0 0 1 2 4 IEC60065 2 10 A 0 1 0 001 1 s 50 Hz 0 0 1 3 5 IEC60065 3 10 A 0 2 0 001 1 s 50 Hz 0 0 1 4 6 IEC60204 1 10 A 0 1 0 001 10 s 50 Hz 0 0 1 5 7 IEC60335 1...

Страница 146: ...test Routine test Main test points A Between the following sections of class I devices requiring grounding 1 Accessible conducting parts 2 Protective shielding 3 Metal barriers B Protective conductor...

Страница 147: ...s 1 0 V maximum Type test Routine test Main test points A Between class 0 and I devices and all accessible conducting parts of class I devices B Protective conductor terminal or protective grounding p...

Страница 148: ...ust not exceed 2 5 V In measurements of resistance do not include the value for the AC power cord s protective grounding wire Type test Routine test Main test points A Between the following sections o...

Страница 149: ...less 2 minutes Rating of the circuit under test is over 30 A to 60 A inclusive 4 minutes Rating of the circuit under test is over 60 A to 100 A inclusive 6 minutes Rating of the circuit under test is...

Страница 150: ...A 8 Appendix TOS6200A 6210...

Страница 151: ...15 DISP2 4 15 Double Action 4 19 E Ending the test 4 15 Excerpts from the safety standards A 4 F FAIL judgment 4 16 4 30 Fail Mode 4 18 Fan 3 6 7 4 Feet 3 3 Firmware version 1 2 Firmware version of p...

Страница 152: ...on fuction 4 36 Protection status 4 36 R RC01 TOS 1 6 1 7 RC02 TOS 1 6 Ready status 4 14 Rear panel 3 6 Recalling a panel memory and program 5 7 Recalling the memory 4 26 Recalling the program 4 28 RE...

Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...

Страница 154: ...KIKUSUI ELECTRONICS CORP http www kikusui co jp en 1 1 3 Higashiyamata Tsuzuki ku Yokohama 224 0023 Japan Tel 81 45 593 7570 Fax 81 45 593 7571 Website...

Отзывы: