1-4
General
TOS6200A/6210
•
Reduced cycle time
A test current is a constant current value set within approximately 100 ms.
Earth continuity tests can be conducted even at 1-second intervals. This enables
the instrument to perform tests in production lines that require reduced cycle
times.
•
Compact and lightweight
A new high-efficiency power supply (achieving a power conversion efficiency of
65%) and large output (TOS6200A: 150 VA, TOS6210: 220 VA) makes possible a
tester that is remarkably compact and lightweight, about half the size and weight
of our conventional testers.
•
Excellent measurement accuracy
The tester is equipped with an ohmmeter of +/-(2% of r 0.003
) that cal-
culates resistance values based on measured current and voltage.
•
Offset canceling function
The tester has an offset canceling function that cancels values for contact resis-
tance at alligator clip connections, the contact resistance of measuring leads in
measurements using two terminals, and other resistance components.
•
Contact check function
The tester is provided with a contact check function that verifies the connection of
the DUT (by current detection) before initiating testing.
•
Voltage judgment function
This function allows either a resistance value or a voltage drop value (voltage
between the SAMPLING terminals) to be selected as the reference value for judg-
ment criteria.
It also allows voltage drop value-based testing to be conducted, which is required
by safety standards such as UL 60950-1 and IEC 60950-1.
•
Simple operations
Tester functions are fast and easily controlled, with an intuitive control method
that's easy to pick up, even for first-time users.
For example, test conditions are set simply by selecting an item displayed on the
LCD with the cursor keys and turning the rotary knob to set/select a value.
Any items you wish to set up can be assigned to the function keys.
CH1: Current waveform
CH1: 50 A/div
20 ms/div
TOS6210
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...