TOS6200A/6210
GPIB and RS-232C
6-41
26
FAILMODE(FMOD)
0|1|OFF|ON
Setting for fail mode.
(To clear a FAIL or PROTECTION condition, press the
STOP switch.)
✗ ✗
27
FAILMODDE?
(FMOD?)
Returns the ON/OFF status for fail mode.
✓ ✓
28
FREQUENCY
(FREQ)
50|60
Hz
Setting for test frequency
✗ ✗
29
FREQUENCY?
(FREQ?)
Returns the value set for test frequency.
✓ ✓
30
FUNCTION(FUN)
0|1|2|3|4
Mode switching
0:MAIN
1:AUTO
2:AUTO EDIT
3:SYSTEM 1
4:OFFSET
✗ ✗
31
FUNCTION?
(FUN?)
Returns the current mode (displayed screen)
✓ ✓
32
IDATA?(IDAT?)
Returns a measured current value.
✓ ✓
33
INVALID?(INV?)
Returns the value of the invalid setting register.
✓ ✓
34
LOWER(LOW)
0.001 to 1.200
(TOS6200A)
0.001 to 0.600
(TOS6210)
0.001
OHM Setting for the lower reference value.
✗ ✗
0|1|OFF|ON
ON/OFF setting for the lower limit judgment function.
35
LOWER?(LOW?)
Returns the value set for the lower reference value.
Returns the ON/OFF status for the lower limit judg-
ment function.
✓ ✓
36
MEASMODE
(MMOD)
NORM|MAX
Setting for the measurement mode.
✗ ✗
37
MEASMODE?
(MMOD?)
Returns the setting status of the measured result indica-
tion.
✓ ✓
38
MEMORY(MEM)
0 to 99
1
Memory number
✗ ✗
20H to 7EH
Memory name (maximum of 12 characters; ASCII
codes """, "'", ",", and "@" cannot be used.)
3.0 to 30.0
0.1
A
Current value
0.001 to 1.200
0.001
OHM Upper reference value
0.001 to 1.200
0.001
OHM Lower reference value
0.3 to 999
0.1, 1
s
Test time
Resolution: “0.1” if the set value is from 0.3 to 99.9
“1” if the set value is from 100 to 999
50|60
Hz
Frequency
0|1|OFF|ON
ON/OFF for the lower reference value
0|1|OFF|ON
ON/OFF for the offset function
0|1|OFF|ON
ON/OFF for the timer function
39
MEMORY?(MEM?)
0 to 99
1
Reads the contents of a corresponding memory and
returns the memory name, currnet value, upper refer-
ence value, lower reference value, test time, frequency
set value, ON/OFF status for the lower limit judgment
function, ON/OFF status for the offset function, and
ON/OFF status for the timer function.
✓ ✓
Header
Data
Function and response data
1
2
Set value
Reso-
lution
Unit
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...