6-14
GPIB and RS-232C
TOS6200A/6210
COMMENT (COM)
Inputs comments or inquires about its contents.
This message is cleared by the *RST command.
Message
Command message:
COM “<DATA>”,“<DATA>”,“<DATA>”
Query message:
COM?
Program data
Set value:
20, maximum number of characters
ASCII code 20H to 7EH
22H (“), 27H (‘), 2CH (,), and 40H (@) cannot be used.
For more information, see A.2, “ASCII Codes 20H to 7EH.”
(Example)
COM "KIKUSUI","TOS6210","!#$%&()*+"
Response
Returns the contents of the comments in response to COM?.
This fills blank entries with spaces so that each line has 20 characters, then returns
this as the result.
(Example) If "KIKUSUI" is entered in the 1st line, no data is entered in the 2nd
line, and "123456789ABCDEFGHIJK" is in the 3rd line, the response
message returns the following result: (
_
= space)
KIKUSUI_ _ _ _ _ _ _ _ _ _ _ _ _,_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _,1
23456789ABCDEFGHIJK
CONTACTCHECK (CCH)
Sets the contact check function to ON/OFF, or inquires about its ON/OFF status.
Message
Command message:
CCH {0|1|OFF|ON}
Query message:
CCH?
Program data
Set value:
0, OFF
Contact check function OFF (default)
1, ON
Contact check function ON
Response
Returns the ON/OFF status of the contact check function in response to CCH?.
(Example) When the contact check function is ON
A value of
1
is returned.
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...