TOS6200A/6210
Basic Operation
4-37
4.10.1 Time Limitation with Respect to Output (OVER HEAT)
The heat radiation capacity at the output section of the tester is designed to be one-
third of the rated output by taking into account size, weight, cost, and other factors.
Use the tester within the limitations given in the table below. If the tester is used in a
way that causes the tester to exceed the limitations, the output section temperature
will rise, which may cause the tester to enter protection status.
If protection status is triggered and the “OVER HEAT” indication blinks on the
LCD, stop testing for approximately the pause time specified in the table below and
then press the STOP key. If the output section temperature has returned to the nor-
mal level, the tester will return to ready status.
Output Time Limitation
Ambient Temp. t (°C)
Test Current I (A)
Pause Time
Max. Test Time
TOS
6200A
At t
40
C
15 < I
30
Equal to or lon-
ger than test time
30 minutes or
less
I
15
Not required
Continuous out-
put available
TOS
6210
40 < I
60
Equal to or lon-
ger than test time
10 minutes or
less
20 < I
40
Equal to or lon-
ger than test time
30 minutes or
less
I
20
Not required
Continuous out-
put available
Test time
A
Pause time
B
Time
CURRENT LOWER UPPER
TIMER
25.0
A
0.100 999
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
OFF
50Hz
UPPER
OVER HEAT
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...