TOS6200A/6210
GPIB and RS-232C
6-9
*STB?
Used to inquire about the contents of the status byte register.
For details on the status byte register, see 6.4, "Registers.”
Message
Query message:
*STB?
Response
Returns the contents of the status byte register to *STB?.
(Example) If Bit 4 of the status byte register is set, “
16
” is returned.
CLR
Clears the all registers exclusive of enable register and sets STOP flag.
The same processing is performed as when the GPIB bus line message DCL or SDC
has been received. Also used to perform the same processing as is performed by a
DCL message from the RS-232C.
Message
Command message:
CLR
DSE
Used to set or reset each bit of the device status enable register.
Also used to inquire about the contents of the device status enable register.
For details on the device status enable register, see 6.4, "Registers.”
Message
Command message:
DSE <NR1>
DSE <HEX>
Query message:
DSE?
Program data
Set value:
0 to 255
(default is 128)
00H to FFH
(default is 80H)
Resolution:
1
(Example)
DSE #H01
Response
Returns the contents of the device status enable register to DSE?.
(Example) If Bit 5 of the device status enable register is set, “
32
” is returned.
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...