6-12
GPIB and RS-232C
TOS6200A/6210
SILENT (SIL)
Used in control via RS-232C to specify whether to return an acknowledgment mes-
sage to a message divided by the response terminator. Also used to inquire about the
set value that specifies whether to return an acknowledgment message.
As an acknowledgment message, either “OK” or “ERROR” is returned.
If an acknowledgment message is to be received, the RS-232C must be set at full
duplex communication.
Full duplex communication: The transmission of data in two directions simultane-
ously. For full duplex settings, see the manual for
your PC.
Message
Command message:
SIL {0|1}
Query message:
SIL?
Program data
Set value:
0
Return an acknowledgement message.
1
Do not return an acknowledgement mes-
sage (default).
Response
The set value for the SILENT message is returned to SIL?.
(Example) If settings are made so that an acknowledgment message will not be
returned, the message “
1
” is returned.
TRM
Used to set the response terminator, then inquire about the set value it. Note that the
GPIB uni-line message “EOI” is effective only in GPIB communication.
Message
Command message:
TRM{0|1|2|3}
Query message:
TRM?
Program data
Set value:
0
CR/LF + EOI (default)
1
LF + EOI
2
EOI
3
CR + EOI
Response
The response terminator currently set is returned to TRM?.
(Example) If the response terminator is set at LF+EOI, the message “
1
” is returned.
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...