8-4
Specifications
TOS6200A/6210
8.2
Interface and Other Functions
Item
Specifications
REMOTE
5-pin DIN connector on the front panel
Function
An option may be connected to provide remote-controlled tester start/stop func-
tions.
Connectable options
Remote controller: RC01-TOS or RC02-TOS
Test probe: LP01-TOS or LP02-TOS
SIGNAL I/O
D-SUB 25-pin connector on the rear panel
For connector pin configuration, see Table 1 (attached).
Input specifi-
cations
High-level input volt-
age
11 V to 15 V
All input signals are controlled in active Low.
The input terminal is pulled up to +12 V with a resistor.
Leaving the input terminals open is equivalent to inputting
a high-level signal.
Low-level input volt-
age
0 V to 4 V
Low-level input cur-
rent
-5 mA maximum
Input time width
5 ms minimum
Output specifi-
cations
Output method
Open collector output (4.5 Vdc to 30 Vdc)
Output withstand
voltage
30 Vdc
Output saturation
voltage
Approximately 1.1 V (at 25
C)
Maximum output cur-
rent
400 mA (TOTAL)
RS-232C
D-SUB 9-pin connector on the rear panel (compliant with EIA-232-D)
Function
All functions except for the POWER switch and KEY LOCK function may be
controlled remotely.
Baud rate
9600, 19200, or 38400 bps
GPIB
Complies with IEEE Std.488-1978
Function
All functions except for the POWER switch and KEY LOCK function may be
controlled remotely.
For the interface function, see Table 2 (attached).
Display
240 x 64 dots LCD, which displays set values, measured values, judgment
results, and other information.
Test function
AUTO
Performs testing (maximum 100 steps) under different test conditions.
MAIN
Performs a single earth continuity test.
Memory function
AUTO
Provides programming of up to 100 different tests, each consisting of a maxi-
mum of 100 steps. (Note that the total number of steps is 500.)
MAIN
100 maximum
Backup battery life
3 years or more (at 25
C)
TEST MODE
PASS HOLD
Allows a PASS judgment to be held for a specified holding time.
MOMENTARY
Allows testing only while the START switch is pressed.
FAIL MODE
Disables cancellation of FAIL or PROTECTION status using a stop signal from
the remote control.
DOUBLE ACTION
Allows testing only when the START switch is pressed within a half-second
after the STOP switch is released.
CONTACT CHECK
Enables the tester to detect a current flowing through the output terminals to
begin testing.
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...