TOS6200A/6210
Basic Operation
4-35
OVER RESI
If a setting meeting the following equation (a setting exceeding the operation range
) is made, the “READY” indication in the upper right of the LCD changes to
blink “OVER RESI,” informing you that testing cannot be conducted under the cur-
rent conditions.
TOS6200A: Upper reference value (voltage value) / test current value >
TOS6210: Upper reference value (voltage value) / test current value > 0.6
In such a case, set the test current value and upper reference value so that the setting
falls within the operation range
. If a resistance value is set to the upper reference
value, OVER RESI check is not made.
4.9.2
Upper Reference Value <= Lower Reference Value
(UP <= LOW)
If a lower reference value more than the upper reference value is set when lower
limit judgment is ON, the “READY” indication in the upper right of the LCD
changes to blink “UP <= LOW,” informing you that testing cannot begin under the
current conditions.
In this case, increase the upper reference value or reduce the lower reference value.
a
CURRENT LOWER UPPER
TIMER
6.0
A
4.50
V
60.0
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
UPPER
50Hz
OVER RESI
a
CURRENT LOWER UPPER
TIMER
25.0
A
1.100 60.0
s
MAIN
MAIN
FREQ
LOWER
ON
OFFSET
OFF
TIMER
ON
50Hz
UP<=LOW
LOWER
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...