TOS6200A/6210
Basic Operation
4-11
4.2.5
Test Time
If the measured value is within the range of the lower reference and upper reference
values after the specified time has elapsed from the start of test, the test result will
be judged PASS, completing the test.
Whether or not to set a test duration is selectable. To set a test duration, set the
TIMER indication to ON in the test duration setup screen. If no test duration is set,
testing will continue until the test result is judged FAIL or you press the STOP
switch.
Timer ON/OFF Settings
To run the test with a specified test duration, set the TIMER indication to ON.
Pressing the SHIFT and F5 keys together allows you to toggle the timer function
ON/OFF. Pressing SHIFT + F5 toggles between ON and OFF.
You can also use the
▲▼
keys to move the cursor to the timer function ON/
OFF indicator. With the cursor at the TIMER indication use the rotary knob to set.
Turning the knob clockwise:
ON
Turning the knob counterclockwise: OFF
Setting the Test Time
You can set a testing duration anywhere in the range 0.3 s to 999 s (resolution of
0.1 s for 0.3 s to 99.9 s and 1s for 100 s to 999 s).
1.
Press the F5 (TIMER) key to move the cursor if the cursor is not below
the timer (you can also move the cursor using the
▲▼
keys).
2.
Use the
keys to move the cursor to the desired digit.
3.
Use the rotary knob to set the test duration.
25.0
A
0.100 60.0
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
READY
50Hz
F1 F2 F3 F4 F5
SHIFT
FREQ
50/60Hz
LOWER
ON/OFF
OFFSET
ON/OFF
TIMER
ON/OFF
UPPER
JUDGE
R/V
▲
▲
CURRENT LOWER UPPER
TIMER
25.0
A
0.100 60.0
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
READY
50Hz
F1 F2 F3 F4 F5
UPPER
▲
▲
▲
▲
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...