6-24
GPIB and RS-232C
TOS6200A/6210
OFFSET (OFF)
Sets ON/OFF for offset cancel function or inquires about ON/OFF status.
Message
Command message:
OFF {0|1|OFF|ON}
Query message:
OFF?
Program data
Set value:
0, OFF
Offset cancel function OFF (default)
1, ON
Offset cancel function ON
Response
Returns the ON/OFF status of the offset cancel function in response to OFF?.
(Example) When offset cancel function is ON
A value of
1
is returned.
TIMER (TIM)
Sets or inquires about the test time and the timer ON/OFF setting.
Message
Command message:
TIM <timer>, {0|1|OFF|ON}
Query message:
TIM?
•
Program data<timer NR2>Test time
Set value:
0.3 to 999
(default is 1.0)
Resolution:
0.3 to 99.9
0.1
100 to 999
1
Unit:
s
•
Program data{0|1|OFF|ON}Timer function’s ON/OFF
Set value:
0, OFF
Timer function OFF (default)
1, ON
Timer function ON
Even when the timer function is set to OFF, you must enter a test time.
In this case, enter any value (Example: maximum value).
(Example)
TIM 999,0
Response
Returns the test time and ON/OFF status for the timer function in response to TIM?.
(Example) When the test time is 60 s and the timer function is ON
A value of
60.0,1
is returned.
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...