TOS6200A/6210
Basic Operation
4-41
4.11 Initialize
• Initializing clears the contents of the programs stored in memory. Before initial-
ization, make sure that the data in memory can be erased safely.
Settings corresponding to the various safety standards are stored in memory. For
information on these settings, see Appendix A.3, “Initial Settings of the Memory.”
Press the POWER switch with the SHIFT key held when starting the tester. This
restores tester settings to factory-set values. (Press the SHIFT key until the charac-
ters "KIKUSUI ELECTRONICS CORP." disappear.)
■
Factory-shipped settings
TOS6200A
TOS6210
Test conditions set up screen (MAIN)
Test current (CURRENT)
3.0 A
6.0 A
Frequency (FREQ)
50 Hz
Lower limit judgment (LOWER)
OFF
Lower reference value (LOWER)
0.001
0.01 V
Upper reference value (UPPER)
0.100
0.60 V
Offset (OFFSET)
OFF
Timer (TIMER)
OFF
Timer set value (TIMER)
1.0 s
Offset screen (OFFSET)
Offset value (OFFSET)
0.000
System screen (SYSTEM)
MEAS MODE
NORM
PASS HOLD
0.2 s
MOMENTARY
OFF
FAIL MODE
OFF
DOUBLE ACTION
OFF
CONTACT CHECK
OFF
BUZZER VOL
4
CONTRAST
6
COMMENT
cleared
Interface screen (INTERFACE)
GPIB ADDRESS
3
SPEED
19200
DATA
8 bit
PARITY
NONE
STOP
2 bit
NOTE
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...