5-6
REMOTE and SIGNAL I/O
TOS6200A/6210
■
Internal configuration
The input and output signal circuits share the same common.
Because the input signal circuit is pulled up to +12 V, opening the input terminals is
equivalent to inputting a high-level signal.
Fig. 5-4
Internal Configuration of the SIGNAL I/O Connector
5.2.2
Starting a Test
To start a test by using the SIGNAL I/O connector, set the ENABLE signal to a low
level first. After a lapse of 10 ms or more from the READY signal has turned to a
low level, set the START signal to a low level for 5 ms or more. The READY signal
turns to a high level after the effective START signal has been detected.
When the ENABLE signal is the low level, the START signal for the SIGNAL I/O
connector and the START input for the REMOTE terminal is enabled. At the same
time, the START switch on the panel are disabled.
Fig. 5-5
START Signal
・
・
・
・
・
・
Output
signal
control
block
Input
signal
control
block
COM
Output signal 1
Output signal 2
Input signal 1
Input signal 2
+12 V
+12 V
READY (OUT)
ENABLE (IN)
START (IN)
10 ms or
more
5 ms or more
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...