TOS6200A/6210
GPIB and RS-232C
6-17
MOMENTARY (MOM)
Sets ON/OFF for the momentary or inquires about ON/OFF status.
• Releasing the START switch when momentary is ON is equivalent to pressing the
STOP switch. If the interval of a step has been set to HOLD in a programmed
automatic test, the program cannot proceed to the next step.
Message
Command message:
MOM {0|1|OFF|ON}
Query message:
MOM?
Program data
Set value:
0, OFF
Momentary OFF (default)
1, ON
Momentary ON
Response
Returns the ON/OFF status of the momentary in response to MOM?.
(Example) When the momentary is ON
A value of
1
is returned.
PASSHOLD (PHOL)
Sets the PASS hold time or inquires about the value set for PASS hold time.
Message
Command message:
PHOL {<NR2>|HOLD}
Query message:
PHOL?
Program data
Set value:
0.2 to 10.0
(default is 0.2)
HOLD
Sets the PASS hold time to HOLD.
Resolution:
0.1
Unit:
s
Response
Returns the set value of PASS hold time in response to PHOL?.
(Example) When the PASS hold time has been set to 2 seconds
A value of
2.0
is returned.
(Example) When the PASS hold time has been set to HOLD
A value of
HOLD
is returned.
NOTE
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...