4-16
Basic Operation
TOS6200A/6210
When the timer is ON
After the time set in the timer elapses, the tester returns a PASS judgment and
reverts to ready status. The measurement result is displayed while "PASS" is dis-
played.
When the timer is OFF
The test continues as long as the measured value meets the conditions set for a PASS
judgment.
To stop the test, press the STOP switch. If the test is terminated by pressing the
STOP switch, no judgment is returned for the measured value, and "PASS" is not
displayed.
■
FAIL judgment
If the lower limit judgment function is ON and the measured value falls outside the
range between the upper and lower reference values, the tester returns a FAIL judg-
ment.
When the lower limit judgment function is OFF, the tester returns a FAIL for the
measured value if the value exceeds the upper reference value.
The "
FAIL" appears in the upper right part of the LCD when the measured value
is below the lower reference value, and the "
FAIL" appears in the same part of the
LCD when the measured value is above the upper reference value. The FAIL LED
indicator also lights, and the buzzer sounds.
To remove a FAIL judgment, press the STOP switch.
The measurement result is displayed until you press the STOP switch.
In case of FAIL, the test duration indicates the elapsed time regardless of the timer
ON/OFF setting.
0.3
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
FAIL
50Hz
3.50
V
25.0
A
0.140
NORM
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...