8-2
Specifications
TOS6200A/6210
8.1
Basic Performance
Item
TOS6200A
TOS6210
Output block
Current setting range *1
3.0 Aac to 31.0 Aac
(With respect to resistance resulting in
output power of the maximum rated
Output or less and an output terminal
voltage of 5.4 V or less)
6.0 Aac to 62.0 Aac
(With respect to resistance resulting in
output power of the maximum rated
Output or less and an output terminal
voltage of 5.4 V or less)
Resolution
0.1 A
Accuracy
(1% of s 0.2A)
(1% of s 0.4A)
Maximum rated output
150 VA (at output terminals)
220 VA (at output terminals)
Distortion factor
2% or less (with respect to 0.1
pure
resistance load of 10 A or greater)
2% or less (with respect to 0.1
pure
resistance load of 20 A or greater)
Frequency
50/60 Hz, sine wave (selectable)
Accuracy
200 ppm
Open terminal voltage
6 Vrms or less
Output method
PWM switching method
Output ammeter
Measurement range
0.0 Aac to 33.0 Aac
0.0 Aac to 66.0 Aac
Resolution
0.1 A
Accuracy
(1% of r 0.2 A)
(1% of r 0.4 A)
Response
Mean value response/rms value display (response time: 200 ms)
Hold function
The current measured at the end of test is held during the PASS or FAIL inteval
Output voltmeter
Measurement range
0.00 Vac to 6.00 Vac
Resolution
0.01 V
Offset cancel function
0.00 V to 5.40 V (Offset ON/OFF function provided)
Accuracy
(1 % of r 0.02 V)
Response
Mean value response/rms value display (response time: 200 ms)
Hold function
The voltage measured at the end of test is held during the PASS or FAIL inteval
Ohmmeter *2
Measurement range
0.001
to 1.200
0.001
to 0.600
Resolution
0.001
Offset cancel function
0.000
to 1.200
(Offset ON/OFF function provided)
0.000
to 0.600
(Offset ON/OFF function provided)
Accuracy
(2 % of r 0.003
)
Hold function
The resistance measured at the end of test is held during the PASS or FAIL inteval
Pass/fail judgment function *3
Resistance value-based judgment
Window comparator system
If a resistance value equal to or greater than the upper reference value is detected,
a FAIL determination is returned.
If a resistance value equal to or less than the lower reference value is detected, a
FAIL determination is returned.
If a resistance value has been judged as FAIL, the tester shuts off the output and
generates a FAIL signal.
If the set time elapses without abnormalities, the tester shuts off the output and
generates a PASS signal.
Setting range for the upper refer-
ence value (UPPER)
0.001
to 1.200
0.001
to 0.600
Setting range for the lower refer-
ence value (LOWER)
0.001
to 1.200
0.001
to 0.600
Resolution
0.001
Judgment accuracy
(2% of UPPER + 0.003
)
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...