Basic Measurements
13.5.4
Average Velocity (Auto Trace)
To measure the Average Velocity of a certain section in Spectral-
Doppler mode using Auto Trace:
1. Select
the
[
Auto Trace
]
item.
It traces the Doppler spectrum automatically and displays the results.
2. Select the sensitivity of the envelope curve
(to eliminate artifacts).
3. Select the Trace Mode channel of the envelope curve (upper, both or lower).
4. If necessary, select the
[
Angle
]
and the
[
Baseline
]
.
A green line appears at the left of the spectrum.
Press the upper trackball key
[
Change
]
to move the line and readjust the start cycle (the line changes to
yellow).Press the right or left trackball key
[
Set
]
to fix the line.
A green line appears at the right of the spectrum. Press the
[
Change
]
key again (line changes to
yellow), move the line to readjust the end cycle and fixate it with
[
Set
]
.
The status area shows the current function of the trackball.
5. Press the right or left trackball key to set the value and finish the
measurement.
Remarks:
•
The measurement results (such as the time and average velocity between two points, VTI,
RI and PI) appear on the monitor screen.
Important note:
The determination of the envelope curve requires a clear and low-noise recording of the Doppler
spectrum. Otherwise the reliability of the displayed measurement results may not be ensured!
Voluson
®
730Pro - Operation Manual
105831 Rev. 0
13-13
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...