CFM Mode
8.3.9 CFM
Map
This function provides selectability of the color coding for an optimization of the display of blood
flow (similar to the post-processing curves with grayscale 2D). It is useful especially with low flow
rates. It may be altered in real time or Freeze mode, respectively.
Velocity display (Display V), velocity-turbulence display (Display V-T) and velocity-power display
(Display V-Pow) each have different color pattern maps available for selection.
Selection of a CFM Map curve:
Switch the
[
C Map
]
key repeatedly to select the desired CFM Map. (1 to 8)
8.3.10 Balance
“Balance” controls the amount of Color displayed over bright echoes and helps confine Color within
the vessel walls. Raising this Balance displays Color on brighter structures. If you see Color on vessel
walls, the Balance is probably set too high. Additionally, wall motion ghosting can be suppressed with
a low Balance setting.
Rotate the
[
Balance
]
digipot and select the desired Balance.
The Balance orientation line is only visible in the color modes.
The line represents to the position of the adjusted gray value in the
gray scale.
Where the balance orientation line is on a gray step, the gray value
will be displayed (only if a color value is present, of course).
For example, if the gray value is higher than 96 while a color value is
present, the gray value will be displayed.
Remarks:
*
Changes are only possible in write mode!
Voluson® 730Pro - Basic User Manual
8-14
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...