Calculations and Patient Reports
14.7.4.8
Heart Rate (HR)
1. Press
the
[
Calc
]
key.
The Cardiac calculations menu appears in the menu area.
2. Select the item Heart Rate
[
HR
]
in the menu.
3. When the vertical line appears on the screen, move it to the first
point to measure using the trackball and the right or left trackball
[
Set
]
key. The second line appears.
4. Select the number of heart rate cycles necessary for measurement with the digipot.
5. Move the second line to the selected number of heart rate cycles to be measured.
Then press the right or the left trackball
[
Set
]
key. The Heart Rate is displayed.
14.7.4.9
MV (Mitral Valve) in M Mode
There are two methods to calculate MV (Mitral Valve) in the M mode.
One method is to calculate all items at a time and the other is to calculate one by one.
•
To measure the Items One by One
(chapter 14.7.4.9.1)
•
To measure all Items at a Time
(chapter 14.7.4.9.2)
14.7.4.9.1
To measure the Items One by One
To measure the items such as D-E, EPSS, E-F Slope, A-C Interval:
1. Press
the
[
Calc
]
key on the control panel.
The Cardiac calculations menu appears in the menu area.
2. Select
[
MV M-Mode
]
and select the appropriate item.
3.
Perform the measurements by using the trackball and the right or left trackball key.
Voluson
®
730Pro - Basic User Manual
14-34
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...