Calculations and Patient Reports
Spectral Doppler Measurements include:
•
Auto Trace
(chapter 14.9.3.1)
•
Manual Trace
(chapter 14.9.3.2)
•
The Measurement of Each Item
(chapter 14.9.3.3)
•
Heart Rate
(chapter 14.9.3.4)
To measure the Heart Rate is also possible in
M mode
.
2D mode Measurements include:
•
To calculate Stenosis %
(chapter 14.9.3.5)
•
To measure Vessel Area / Distance
(chapter 14.9.3.6)
14.9.3.1 Auto
Trace
To perform the measurement using Auto Trace:
1. Press
the
[
Calc
]
key on the control panel.
The Vascular Calculations menu appears in the menu area.
2. Select the appropriate “Measure Group” and select
[
Auto Trace
]
.
The Doppler spectrum is traced automatically and the results are
displayed.
3. Select the sensitivity of the envelope curve
(to eliminate artifacts).
4. Select the Trace Mode channel of the envelope
curve. (upper, both or lower)
A green line appears at the left of the spectrum.
Press the upper trackball key
[
Change
]
to move the line and readjust
the start cycle (the line changes to yellow).Press the right or left
trackball key
[
Set
]
to fix the line.
A green line appears at the right of the spectrum. Press the
[
Change
]
key again (line changes to yellow), move the line to readjust the end
cycle and fixate it with
[
Set
]
.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
14-45
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...