Utilities
12.1 Gray Chroma Map
A gray map determines the displayed brightness of an echo in relationship to its amplitude.
A Chroma Map determines the displayed Color of an echo in relationship to its amplitude.
Depending on individual requirements a "harder" or "softer" image can be obtained with this function
and can be adjusted in Read- and in Write-Mode (post-processing). The displayed gray scale wedge
corresponds to the adjusted gray map curve.
Different gray map curves can be related to a 2D image, M image, D image and 3D Volume.
•
To select a 2D Gray map review:
2D Gray Map
(chapter 12.1.1)
•
To select a 3D Gray map review:
3D Gray Map
(chapter 12.1.2)
•
To select a Chroma map review:
Chroma Map
(chapter 12.1.4)
12.1.1
2D Gray Map
This defines the relation between echo amplitude (input) and brightness (output) in a look-up table.
In total 9 predefined and 3 user Gray Maps are available. You can select a certain map independently
for each one of the 2D-, D-, M- imaging modes (for example: Map 5 for 2D Imaging and Map 2 for
M-Imaging etc.). For 3D Gray Map selection review: 3D Gray Map.
Press the
[
Utilities
]
key regardless which mode is active.
Operation:
1. Select
[
Gray/Chroma
]
. The “Gray Chroma” menu will appear.
2. Select the Mode where to change the Gray Map curve.
Key bright: the Gray map selection assigned to this mode
3. Select the Gray Predefined or the Gray User
curve with the flip switch.
Return to the last active menu.
To create your own Gray Map
review:
Gray
Edit Menu
(chapter 12.1.3)
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
12-3
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...