Volume Mode
Definition of the Surface Geometry
The
Surface Geometry
is defined by rotation of an image plane about a fixed axis (main contour axis)
and the definition of 2D contours in each plane. The 2D contours can be defined manually or by a
sphere. The rotation step for each contour plane is selectable in the range from 6 to 30 degrees (30 - 6
contour planes).
The
Surface Geometry
is defined by 3D triangularization of the 2D contours, meaning each point of
the 2D contour in plane N is connected via a triangle mesh to corresponding points in plane N-1 and
plane N+1.
Definition of a Shell Contour (Geometry)
The basic idea of a
Shell Contour
is, to define a “thickness”' of the “reference”' surface geometry.
The “'parallel” contours shown in the image define the
“parallel” surface geometry (describing the shell).
The “parallel” contours are either defined symmetrically to
the reference contour or limited to one direction, inside or
outside.
The
Shell Geometry
consists of one outside and one inside
surface and therefore it is possible to distinguish between
points enclosed by the shell geometry and points outside of
it. A
Shell Contour
represents all points enclosed by the
inner and outer surface geometry.
If no
Shell Contour
is defined explicitly, the
Shell Geometry
consists of the reference surface (outside surface) and an
inner point (the inside surface being degenerated).
Display of a Shell Geometry (contour rendering)
The shell geometry can be visualized as “Skin” or “Wire Mesh”.
review:
Render Mode and Display of the Shell Geometry
(chapter 11.6.4)
The image shows the different visualization techniques. VOCAL shows a surface Mesh.
.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
11-65
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...