2D Mode
5.19.2 Line
Filter
With “Line Filter” especially the lateral resolution can be optimized with this innovative correlation
algorithm. With the new process, the signals of the neighboring pulses are less weighted for the
display of the actual pulse which considerably improves the detail resolution and signal-to-noise ratio.
Three steps are provided: off, low, high
The line filter state is displayed in the Image Info area on the screen.
high: filtering of three lines (25/50/25%)
low:
filtering of two lines (12,5/75/12,5%)
off: no
filtering
Remarks:
*
Changes are only possible in write mode. (Control has no function in read mode.)
5.19.3 Enhance
With the “Enhance” function the echo information is digitally processed such that certain existing
information becomes easily visible for the eye (e.g., adjacent media layers). Due to the Enhance
function a finer, sharper impression of the image is produced.
Six steps are possible: 0, 1, 2, 3, 4, 5
The Enhance state is displayed in the Image Info area on the screen.
Remarks:
*
Change are only possible in write mode. (Control has no function in read mode.)
5.19.4 Dynamic
Control
“Dynamic Control” allows you to enhance a part of the grayscale to make it easier to display
pathology. You can select between twelve different Dynamic control curves.
The dynamic control is displayed in the Image Info area on the screen.
Dynamic Control: C1 to C12
Voluson
®
730Pro - Basic User Manual
5-20
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...