2D Mode
5.6
2D Automatic Optimization
This function will optimize the contrast resolution according to the histogram of the scan area.
The shape of the ROI will depend on the probe, scan depth and scan angle.
The primary result is a value for the left and right endpoint of the actual histogram.
Pressing the
[
auto
]
key causes automatic optimization of the gray scale to enhance the
contrast resolution.
When the key is pressed again, the optimization according to the histogram will be
updated and remain active.
Press the
[
auto
]
key twice to switch off the Automatic Optimization of the 2D mode image.
Remarks:
*
When the Automatic Optimization function is active, the
[
auto
]
key is brightly illuminated.
*
In CFM, PW, CW, PD, 3D/4D the optimized settings for the 2D image are kept, but the
[
auto
]
function is disabled.
5.7 Transmit
Power
The
[
Transmit Power
]
flip switch governs the acoustic output of the transducer. It shall be set to the
minimum value which still allows well evaluable information. Always keep the power level and the
exposure time
A
S
L
OW
A
S
R
EASONABLY
A
CHIEVABLE.
Use the
[
Power
]
flip switch to adjust the amount of acoustic output.
Remarks:
*
The actual transmit power value is displayed in the monitor’s Image Information area.
*
The maximum possible acoustic output can be reduced by this control function, if certain values
of Mechanical and Thermal Indices are exceeded.
*
The adjustment of transmit power output changes also the actual output adjustment of all other
modes.
Voluson
®
730Pro - Basic User Manual
5-6
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...