2D Mode
5.2
2D Mode Depth
With this function the depth range of the ultrasound image for the region of interest is adjusted. The
number of image lines and the frame rate are automatically optimized. Changing of the depth is only
possible in real time (write mode).
When the
[
Depth
]
control is flipped downwards, the depth range of the 2D image is
enhanced and the display size of the image is reduced to view the entire depth range.
When the
[
Depth
]
control is flipped upwards, the depth range of the 2D image is
reduced and the display size of the image is magnified.
When you
change the Depth, the 2D image display, the depth scale, acoustic power indices
(MI, TIS, TIB, TIC), frame rate, and focal depth will change accordingly.
Remarks:
*
The maximum and minimum depth depend on the selected probe.
The actual depth in [cm] is displayed in the Information Header.
Read
mode:
*
The displayed 2D image is positioned again on the monitor without change in the depth range.
5.3
2D Image Angle
Use the
[
Angle
]
digipot to select a part of interest of the 2D image. The advantage of the decreased
field-of-view is an increased 2D frame rate due to the smaller sector width.
→
Turn clockwise to increase the image width.
←
Turn counterclockwise to decrease the image width.
Remarks:
*
If the selected probe has the possibility to adjust the 2D angle, the digipot control will show its
value.
*
Digipot display:
curved array probe: Angle [degree]
Voluson
®
730Pro - Basic User Manual
5-4
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...