Safety
The following results are contained in the AIUM report too:
•
Other acting mechanisms
- no other acting mechanisms for bioeffects are known, that are caused
by exposure to diagnostic ultrasound.
•
Epidemiology
- in more than 25 years of extensive clinical ultrasound application no harmful
effects due to exposure to diagnostic ultrasound have become known.
•
Bioeffects in vivo with mammals
- no significant bioeffects in vivo independently confirmed could
be determined under the following conditions:
•
Unfocussed exposure with an SPTA intensity of less than 100 mW/cm2 in an exposure period of
less than 500 seconds (measured in water).
•
Focused exposure with an SPTA intensity of less than 1 W/cm2 in an exposure period of less than
50 seconds (measured in water).
•
Product from intensity (measured in water) and time of exposure less than 50 joule/cm2.
2.13.2
Intensities Measured in Water and Recalculated In Situ
All intensity parameters are determined by measurement in water. As water does not absorb the
acoustic energy, these measurements in water represent the most unfavorable value. In biological
tissue however the acoustic intensity is absorbed. The "real" value in a given position depends on the
amount and type of tissue through which the ultrasound beam passes and on the ultrasound frequency.
The value in tissue (
in situ
) can be approximately determined with the following formula:
in situ = water [ e -(0.23dlf) ]
Whereas:
in situ = value in situ
water = value in water
e =
2.7183
d =
attenuation
coefficient
l
= distance from skin surface to measuring depth (cm)
f
= mean frequency of combination probe/system/operation
mode (MHz)
tissue d (dB/cm/MHz)
brain
0.53
heart
0.66
kidney 0.79
liver
0.43
muscle 0.55
As the ultrasound generally crosses tissue layers of different thickness and different types on its way
through the body during an examination, it is very difficult to estimate the real intensity
in situ
.
For reports generally an impedance coefficient of 0.3 dB/cm/MHz is assumed. The value in situ, that
is generally indicated in reports is calculated according to the following formula:
in situ (recalculated ) = water [e -(0.69lf) ]
As this value must not be considered as the real intensity in situ, the term "recalculated" is used
hereunder.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
2-11
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...