M Mode
6.2.8
M Gain Control
With the
[
Gain
]
control the overall brightness of the M mode trace can be adjusted. The adjustment of
the Gain control determines the amount of amplification applied to the received echoes.
All received echoes are amplified with the same gain value regardless of the scan depth.
The M Gain function influences the M trace only.
[
M
-Mode
]
key:
rotate it to adjust the sensitivity (brightness) of the entire image.
When the GAIN control is turned clockwise the entire image gets brighter.
When the GAIN control is turned counterclockwise the entire image gets less bright.
Remarks:
*
GAIN range: -15dB to 15dB. The actual gain value is displayed on the screen [GN ...]
*
Changing the M Gain is only possible with active M mode (write mode).
6.2.9 M
Cineloop
Several 2D image frames and M mode trace information can be recalled.
When freezing, a certain time frame (M information of the last examination sequence) is stored in a
loop memory. The sequence can be reviewed second by second.
Display:
Cine
for 2D images
or
Loop
for M trace on the monitor (status area)
min
length:
60
seconds
Operation:
1.
Freeze
the
image.
After freezing the trackball is active for the M mode trace loop / 2D cine.
2.
The upper trackball key changes from the M loop to the 2D cine, and back.
The active cine is displayed in the trackball graphic on the monitor:
Cine/
Loop xxx
or
Cine xxx/
Loop
3.
Use the trackball to recall the stored sequence by moving it horizontally.
Voluson
®
730Pro - Basic User Manual
6-6
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...