21.2.2
Power Supply (rear side)
F1
Mains Switch:
Thermal Circuitbreaker
T 16 Ampere /230V reusable
Manufacturer:
E-T-A
â
Type: 3120-F521-P7TI-W01A 16A
Mains IN
Mains voltage according to type plate
Modification of voltage setting only by authorized service person!
Possible mains voltages: 100V, 115V, 130V, 230V, 240V, AC only
F4, F5, F6, F7:
T 10 Ampere /250V fuse Type: H (high breaking capacity)
Manufacturer: Schurter order Nr. 0001.2534 Type SPT
F2
On-switch for auxiliary devices:
Thermal
Circuitbreaker T 2,5 Ampere / 230V reusable
T 5 Ampere / 115 V reusable
Manufacturer: E-T-A
â
Type: 3120-F521-P7TI-W01A-X017 2.5/5A
Table for the Fuses used.
Potential equilibrium connection (rear panel)
Protective earth connection
Information plate for the adjusted voltage setting.
Modification of voltage setting only by authorized service person!
Possible mains voltages: 100V, 115V, 130V, 230V, 240V, AC only
Connector for Foot Switch.
review also:
Important Instructions for Safety
(chapter 2.1)
.
To adjust the Foot Switch review: System Setup:
Peripherals
(chapter 17.3.3)
.
Voluson
®
730Pro - Basic User Manual
21-6
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...