Probes and Biopsy
20.8 Probe
Usage
For details on connecting, activating, deactivating, disconnecting, transporting and storing the probes,
refer to:
Transducer Connection
(chapter 4.4)
and
Probe/Program Selection
(chapter 4.5).
20.9 Care and Maintenance
20.9.1 Inspecting
Probes
After each use, inspect the probe's lens, cable, and casing. Look for any damage that would allow
liquid to enter the probe. If any damage is found, do not use the probe until it has been inspected and
repaired/replaced by a GE Medical Systems - Kretztechnik Service Representative.
NOTE:
Keep a log of all probe maintenance, along with a picture of any probe malfunction.
20.9.2 Environmental
Requirements
Probes should be operated, stored, or transported within the parameters outlined below.
Ensure that the probe face temperature does not exceed the normal
operation temperature range.
Probe Environmental Requirements
Operational Storage Transport
Temperature
+10° to +40°
C
+50° to +104° F
-10° to
+60°
C
+14° to +140° F
-40° to
60°
C
-40° to +140° F
Humidity
Max. 90%
non-condensing
Max. 90%
non-condensing
Max 90%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Voluson
®
730Pro - Basic User Manual
20-8
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...