MAN ECG preamplifier
22.5 ECG
Display
This function inserts an ECG line in the display of the image.
Condition: ECG module (ECG preamplifier) is connected to the system.
Operation:
1. Press
the
[
ECG
]
key on the control panel to switch on/off the ECG line.
The status area shows additional functions for the ECG Display.
2. Adjust the transmission gain of the ECG preamplifier signal (0, 1, 2, 3).
3. Select ECG velocity (0, 1, 2, 3).
4. Set the vertical position on the monitor.
5. Adjust ECG amplitude ( 0 to 100 in 10 steps).
6. Return to the main menu.
The ECG function remains active.
7. Freeze the image.
The most recent information is always on the right edge of the image.
When moving the trackball a indicator (small vertical line) is inserted on the
ECG curve and indicates the temporal position of the 2D image in relation to the
recorded ECG-line. In this manner, e.g., diastolic or systolic phase of the 2D mode
image can be set (without ECG trigger).
Voluson
®
730Pro - Operation Manual
105831 Rev. 0
22-5
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...