M Mode
6.3.2 Enhance
With the “Enhance” function the echo information is digitally processed such that certain existing
information becomes easily visible (e.g., adjacent media layers). Due to the Enhance function a finer,
sharper impression of the image is produced.
The Enhance state is displayed in the Image Info area on the screen.
Six steps are possible: 0, 1, 2, 3, 4, 5
Remark:
Changes are only possible in write mode. (Control has no function in read mode.)
6.3.3 Dynamic
Control
“Dynamic Control” allows you to enhance a part of interest of the grayscale to make it easier to
display pathology. You can select between twelve different Dynamic control curves.
The dynamic control is displayed in the Image Info area on the screen.
Dynamic Control:
C1 to C12
Remarks:
*
The appearance of the gray values depends also on the selected gray map.
To select a M mode gray map review chapter:
Gray Chroma Map
(chapter 12.1)
*
Changes are only possible in write mode. (Control has no function in read mode.)
6.3.4 Reject
The “Reject” function determines the amplitude threshold above which the ultrasound echoes
are displayed on the screen (suppression of smaller echoes). Display (control range) 0 to 255.
The Reject state is displayed in the Image Info area on the screen.
Reject range max.:
255
Reject range min.:
0
step size:
5
Remark:
Changes are only possible in write mode. (Control has no function in read mode.)
6.3.5
Exit the M Submenu
Press the
[
Exit
]
key on the control panel to exit the M Submenu.
Voluson
®
730Pro - Basic User Manual
6-8
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...