Spectral Doppler Mode
7.1.2.4
PW Automatic Optimization
This function will optimize the following settings:
PRF:
automatic detection of highest flow velocities and adjustment of the velocity scale
Baseline:
will be shifted so that the flow spectrum is centered
Pressing the
[
auto
]
key enables automatic optimization of the PRF and Baseline.
When the key is pressed again, the optimization will be updated.
Press the
[
auto
]
key
twice to switch off
the Automatic Optimization in PW mode.
Remarks:
*
When the Automatic Optimization function is active, the
[
auto
]
key is brightly illuminated.
*
The PRF and Baseline can always be changed manually!
7.1.2.5 Sweep
Speed
The “Speed” control allows selection of variable sweep speeds. The faster sweep speed may be useful
to analyze flow curves. For example, to calculate a mean pressure gradient then it will be much easier
to do this on a fast speed trace than on a low speed trace.
The
[
Speed
]
key is found in the status area; review:
PW Main Menu
(chapter 7.1.1)
By touching
⇧
or
⇩
, four different sweep speeds can be selected.
1 = 3.5 cm/s
2 = 5.0 cm/s
3 = 7.5 cm/s
4 = 10.0 cm/s (in relation to the system’s monitor)
Remarks:
Changing the speed is only possible in write mode.
7.1.2.6 Audio
Signal
This flip switch changes the volume of the audio signal derived from the PW spectrum.
↑
both
loudspeakers
louder
↓
both
loudspeakers
lower
The volume is adjustable between 0 and 96 dB .
Voluson
®
730Pro - Basic User Manual
7-6
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...