Probes and Biopsy
20.10.4 Mechanical Hazards
A defective probe or excessive force can cause patient injury or probe damage:
•
Observe depth markings and do not apply excessive force when inserting or
manipulating intracavitary probes.
•
Inspect probes for sharp edges or rough surfaces that could injure sensitive
tissue.
20.11 Special Handling Instructions
20.11.1 Using Protective Sheaths
Protective barriers may be required to minimize disease transmission. Probe
sheaths are available for use with all clinical situations where infection is a
concern. Use of legally marketed, sterile probe sheaths is strongly
recommended for intracavitary and intraoperative procedures. Use of legally
marketed, sterile, pyrogen free probe sheaths is REQUIRED for neurological
intra-operative procedures.
Instructions:
Custom-made sheaths are available for each probe. Each
probe sheath kit consists of a flexible sheath used to cover the probe and
cable and elastic bands used to secure the sheath.
Sterile probe sheaths are supplied as part of biopsy kits for those probes
intended for use in biopsy procedures. In addition to the sheath and elastic
bands, there are associated accessories for performing a biopsy procedure
which are included in the kit.
Refer to the biopsy instructions for the specific probes in the Discussion
section of this chapter for further information.
Reordering:
To reorder sheaths, refer to Ultrasound Probe and Cord Sheath
Sets (continued).
Devices containing latex may cause severe allergic reaction in latex sensitive
individuals.
Refer to FDA's March 29, 1991 Medical Alert on latex products.
DO NOT
use pre-lubricated condoms as a sheath.
In some cases, they may damage the probe. Lubricants in these condoms may
not be compatible with probe construction.
DO NOT
use an expired probe sheath.
Before using probe sheaths, verify whether the term of validity has expired.
Voluson
®
730Pro - Basic User Manual
20-10
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...