Calculations and Patient Reports
14.3.4.4.2 Manual
Trace
1. After obtaining a feasible Doppler spectrum, press the
[
Calc
]
key on the control panel.
The OB Calculations menu appears in the menu area.
2. Select the appropriate item and select
[
Manual Trace
]
.
3. The cursor appears on the Doppler spectrum. Move the cursor to the start point of the period and
press the right or left trackball key
[
Set
]
.
Note:
To readjust the traced line press the upper trackball key
[
Undo
]
repeatedly.
4. Trace to the end of the period and press the
[
Set
]
key once more.
Remarks:
•
Depending on the setting in the Measurement Setup, the envelope curve will be performed with a
continuous trace line or by setting points.
To select the Doppler Manual Trace Method review:
General Measurement Setup
(chapter 18.3.1)
14.3.4.4.3
Measurement of Each Item
1. After obtaining an appropriate image, press the
[
Calc
]
key.
2. Select the appropriate item (PSV, EDV).
3. The cursor appears on the screen. Perform the measurement using the right or left trackball key.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
14-15
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...