CFM Mode
8.3.6 Baseline
The CFM baseline shift can be used to prevent aliasing in one flow direction similar to the PW
Doppler baseline shift. Shifting the CFM baseline enlarges the velocity range in one direction.
The zero line of the color bar is also shifted.
There are
8 Steps in each direction.
By step 8 you see only the color wedge in one direction (maximum velocity), the other
direction is 0 (kHz, cm/s , m/s).
Only possible in Display Mode V, V-T and V-Pow
The maximum value and minimum value of the flow velocities are displayed at the upper and lower
edge of the color wedge.
8.3.7 Ensemble
This function controls the number of pulses for one displayed Color-Doppler line. Since several pulses
are to be evaluated for display of a result, the color display quality increases with the number of
evaluated pulses.
Switch the
[
Ensemble
]
key and select the number of pulses per color line.
max. value:
31
min. value:
7
step size:
1
Switch up or down to increase or decrease the number of pulses per color line.
With increasing CFM Ensemble the frame rate decreases.
Remark:
Changes are only possible in write mode!
8.3.8 Line
Density
This function determines the line density within the CFM box. The lower the line density, the greater
the distance between lines and the size of the color pixels.
Switch the
[
Line D.
]
key and adjust the line density.
Switch up or down to increase or decrease the line density.
max. value:
9
min. value:
1
step size:
1
Remark:
Changes are only possible in write mode.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
8-13
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...